Performance comparison of a state-of-the-art neuro-SPET scanner and a dedicated neuro-PET scanner.

D L Bailey1, F Zito, M C Gilardi, A R Savi, F Fazio, T Jones.   

Abstract

The physical performances of two current state-of-the-art scanners dedicated to functional imaging of the brain, one a single-photon emission tomography (SPET) scanner and the other a positron emission tomography (PET) scanner, have been compared under identical conditions. The aim of the study was to compare the capabilities of the devices under conditions resembling the routine clinical environment, as well as to consider other issues such as radiation burden for some common investigations. Both systems have slightly less than 11-cm axial fields of view. The PET system can be operated in a septa-less (3D) mode as well as conventionally with septa (2D). The spatial resolution of both devices was less than 8 mm in all dimensions in scattering media. On average, the PET scanner's resolution was approximately 10%-15% better than the SPET system. Energy resolution on the SPET system was superior due the scintillator used [NaI(Tl)]. Sensitivity in air with a line source on the PET system was found to be approximately 150 times greater in 3D and approximately 25 times greater in 2D than with the SPET system. A normal subject was studied on each system in an attempt to obtain the highest quality data possible for a subjective comparison. It is clear that, while PET retains the advantages of more desirable radiopharmaceuticals and higher sensitivity, the quality obtainable from SPET devices has improved markedly. SPET may prove as useful for many clinical investigations.

Mesh:

Year:  1994        PMID: 8062841     DOI: 10.1007/bf00171411

Source DB:  PubMed          Journal:  Eur J Nucl Med        ISSN: 0340-6997


  15 in total

1.  The radiation dosimetry of 99Tcm-exametazime.

Authors:  R G Soundy; D A Tyrrell; R D Pickett; M Stabin
Journal:  Nucl Med Commun       Date:  1990-11       Impact factor: 1.690

Review 2.  Quantitative SPECT of uptake of monoclonal antibodies.

Authors:  G L DeNardo; D J Macey; S J DeNardo; C G Zhang; T R Custer
Journal:  Semin Nucl Med       Date:  1989-01       Impact factor: 4.446

3.  Deadtime characteristics of Anger cameras.

Authors:  J A Sorenson
Journal:  J Nucl Med       Date:  1975-04       Impact factor: 10.057

4.  Noise equivalent count measurements in a neuro-PET scanner with retractable septa.

Authors:  D L Bailey; T Jones; T J Spinks; M C Gilardi; D W Townsend
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

5.  Fully three-dimensional reconstruction for a PET camera with retractable septa.

Authors:  D W Townsend; A Geissbuhler; M Defrise; E J Hoffman; T J Spinks; D L Bailey; M C Gilardi; T Jones
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

6.  Symbiotic developments in PET and SPECT to quantify and display myocardial tomography.

Authors:  E V Garcia; S J Cullom; J R Galt
Journal:  J Nucl Med       Date:  1991-01       Impact factor: 10.057

Review 7.  PET and SPECT to achieve diagnostic impact.

Authors:  U Buell
Journal:  Nuklearmedizin       Date:  1990-03       Impact factor: 1.379

8.  Improved SPECT using simultaneous emission and transmission tomography.

Authors:  D L Bailey; B F Hutton; P J Walker
Journal:  J Nucl Med       Date:  1987-05       Impact factor: 10.057

9.  An electronically collimated gamma camera for single photon emission computed tomography. Part I: Theoretical considerations and design criteria.

Authors:  M Singh
Journal:  Med Phys       Date:  1983 Jul-Aug       Impact factor: 4.071

10.  Correction for deadtime losses in a gamma camera/data analysis system.

Authors:  K Cranley; R Millar; T K Bell
Journal:  Eur J Nucl Med       Date:  1980-08
View more
  5 in total

1.  The relevance of interictal rCBF brain SPECT in temporal lobe epilepsy: diagnostical value and effects of spatial resolution.

Authors:  C Menzel; A Hufnagel; F Grünwald; L Pavics; K Reichmann; C E Elger; H J Biersack
Journal:  Ann Nucl Med       Date:  1995-11       Impact factor: 2.668

Review 2.  The role of positron emission tomography within the spectrum of medical imaging.

Authors:  T Jones
Journal:  Eur J Nucl Med       Date:  1996-02

3.  A method for calibrating three-dimensional positron emission tomography without scatter correction.

Authors:  D L Bailey; T Jones
Journal:  Eur J Nucl Med       Date:  1997-06

4.  Single-photon emission tomographic quantification in spherical objects: effects of object size and background.

Authors:  F Zito; M C Gilardi; P Magnani; F Fazio
Journal:  Eur J Nucl Med       Date:  1996-03

5.  In vivo follow-up of rat tumor models with 2-deoxy-2-[F-18]fluoro-D-glucose/dual-head coincidence gamma camera imaging.

Authors:  Jacques Monteil; Aurélie Dutour; Barbara Akla; Thierry Chianéa; Valérie Le Brun; Laurent Grossin; François Paraf; Yolande Petegnief; Jean-Claude Vandroux; Michel Rigaud; Franck G Sturtz
Journal:  Mol Imaging Biol       Date:  2005 May-Jun       Impact factor: 3.488

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.