Effects of disposable or interchangeable positive end-expiratory pressure valves on work of breathing during the application of continuous positive airway pressure.

R M Kacmarek1, H Mang, N Barker, M C Cycyk-Chapman.   

Abstract

OBJECTIVE: To determine which of a series of disposable or interchangeable positive end-expiratory pressure (PEEP) devices functions with the least imposition of inspiratory and expiratory work during continuous positive airway pressure.
DESIGN: Prospective laboratory evaluation performed on a lung model.
SETTING: Research laboratory at a university medical center.
INTERVENTIONS: A spontaneously breathing lung model, created from a training test lung and a volume ventilator, were used to simulate a patient spontaneously breathing at a tidal volume of 0.4 L, peak inspiratory flow of 40 L/min, an inspiration/expiration ratio of 1:2, and a respiratory rate of 20 breaths/min. Ten PEEP valves attached to a continuous high-flow system were evaluated.
MEASUREMENTS AND MAIN RESULTS: All of the PEEP valves studied imposed high levels of both inspiratory and expiratory work of breathing. The BE-171 and BE-142 valves (Instrumentation Industries) imposed the least amount of inspiratory work. In general, imposed inspiratory work accounted for approximately 70% to 80% of total imposed work of breathing.
CONCLUSIONS: All of the disposable/interchangeable PEEP valves that were studied imposed a considerable amount of both inspiratory and expiratory work, even when the continuous flow provided exceeded the peak inspiratory flow demands of the lung model. The primary reason for the high imposed work levels is the high gas-flow resistance of all of the valves studied.

Entities:  

Mesh:

Year:  1994        PMID: 8045140     DOI: 10.1097/00003246-199408000-00004

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  2 in total

1.  Comparison of two flow generators with a noninvasive ventilator to deliver continuous positive airway pressure: a test lung study.

Authors:  Carolina Fu; Pedro Caruso; Jeanette Janaina Jaber Lucatto; Guilherme Pinto de Paula Schettino; Rogério de Souza; Carlos Roberto Ribeiro Carvalho
Journal:  Intensive Care Med       Date:  2005-10-13       Impact factor: 17.440

2.  Noninvasive positive pressure ventilation delivered by helmet vs. standard face mask.

Authors:  Davide Chiumello; Paolo Pelosi; Eleonora Carlesso; Paolo Severgnini; Michele Aspesi; Chiara Gamberoni; Massimo Antonelli; Giorgio Conti; Maurizio Chiaranda; Luciano Gattinoni
Journal:  Intensive Care Med       Date:  2003-06-12       Impact factor: 17.440

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.