Literature DB >> 16222519

Comparison of two flow generators with a noninvasive ventilator to deliver continuous positive airway pressure: a test lung study.

Carolina Fu1, Pedro Caruso, Jeanette Janaina Jaber Lucatto, Guilherme Pinto de Paula Schettino, Rogério de Souza, Carlos Roberto Ribeiro Carvalho.   

Abstract

OBJECTIVE: To compare the performance of two continuous flow generators with a ventilator designed for noninvasive positive pressure ventilation (NPPV) to deliver continuous positive airway pressure (CPAP). The performance of flow generators using different oxygen pressure supplies was also compared. DESIGN AND
SETTING: Experimental study using a mechanical lung model in a university research laboratory. MEASUREMENTS: Two flow generators supplied at 100, 200, and 300 kPa and an NPPV ventilator were compared at CPAP of 5, 10, and 15 cmH2O in: (a) area under the adjusted CPAP level during inspiration, (b) capacity to attain the preset CPAP, and (c) tidal volume.
RESULTS: The NPPV ventilator attained the preset CPAP better than flow generators, but its area under adjusted CPAP was similar to or higher than that of flow generators when these were adjusted to their better pressure supply. Both flow generators had better performance with an output flow around 100 l/min, which was achieved at 100 kPa with one flow generator and 300 with the other. Flow generators and the NPPV ventilator generated similar tidal volumes.
CONCLUSIONS: Flow generators performance showed large variations among different devices and oxygen pressure supplies. Adjusted to their better pressure supply, flow generators had a similar or better capacity to maintain the CPAP level, but the NPPV ventilator was more reliable to attain the preset CPAP. Flow generators could be an alternative to provide CPAP in low-income areas, usually with scarce medical equipment availability.

Entities:  

Mesh:

Year:  2005        PMID: 16222519     DOI: 10.1007/s00134-005-2795-x

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  18 in total

Review 1.  International Consensus Conferences in Intensive Care Medicine: non-invasive positive pressure ventilation in acute respiratory failure. Organised jointly by the American Thoracic Society, the European Respiratory Society, the European Society of Intensive Care Medicine, and the Société de Réanimation de Langue Française, and approved by the ATS Board of Directors, December 2000.

Authors:  T W Evans
Journal:  Intensive Care Med       Date:  2001-01       Impact factor: 17.440

2.  Mask mechanics and leak dynamics during noninvasive pressure support ventilation: a bench study.

Authors:  G P Schettino; M R Tucci; R Sousa; C S Valente Barbas; M B Passos Amato; C R Carvalho
Journal:  Intensive Care Med       Date:  2001-11-10       Impact factor: 17.440

3.  Flow resistance of expiratory positive-pressure systems.

Authors:  M R Pinsky; D Hrehocik; J A Culpepper; J V Snyder
Journal:  Chest       Date:  1988-10       Impact factor: 9.410

4.  A continuous flow intermittent mandatory ventilation with continuous positive airway pressure circuit with high-compliance reservoir bag.

Authors:  A Braschi; G Iotti; A Locatelli; G Bellinzona
Journal:  Crit Care Med       Date:  1987-10       Impact factor: 7.598

5.  Efficacy of noninvasive CPAP in COPD with acute respiratory failure.

Authors:  P Goldberg; H Reissmann; F Maltais; M Ranieri; S B Gottfried
Journal:  Eur Respir J       Date:  1995-11       Impact factor: 16.671

6.  The Downs' adjustable flow generator.

Authors:  G C Fisher
Journal:  Anaesthesia       Date:  1988-09       Impact factor: 6.955

7.  Effects of disposable or interchangeable positive end-expiratory pressure valves on work of breathing during the application of continuous positive airway pressure.

Authors:  R M Kacmarek; H Mang; N Barker; M C Cycyk-Chapman
Journal:  Crit Care Med       Date:  1994-08       Impact factor: 7.598

8.  Position of exhalation port and mask design affect CO2 rebreathing during noninvasive positive pressure ventilation.

Authors:  Guilherme P P Schettino; Sunisa Chatmongkolchart; Dean R Hess; Robert M Kacmarek
Journal:  Crit Care Med       Date:  2003-08       Impact factor: 7.598

9.  CO2 rebreathing during BiPAP ventilatory assistance.

Authors:  G T Ferguson; M Gilmartin
Journal:  Am J Respir Crit Care Med       Date:  1995-04       Impact factor: 21.405

10.  Noninvasive positive pressure ventilation delivered by helmet vs. standard face mask.

Authors:  Davide Chiumello; Paolo Pelosi; Eleonora Carlesso; Paolo Severgnini; Michele Aspesi; Chiara Gamberoni; Massimo Antonelli; Giorgio Conti; Maurizio Chiaranda; Luciano Gattinoni
Journal:  Intensive Care Med       Date:  2003-06-12       Impact factor: 17.440

View more
  2 in total

1.  Impact of continuous positive airway pressure (CPAP) on the respiratory capacity of chronic kidney disease patients under hemodialysis treatment.

Authors:  Vivian Bertoni Xavier; Renata Spósito Roxo; Luiz Antônio Miorin; Vera Lúcia Dos Santos Alves; Yvoty Alves Dos Santos Sens
Journal:  Int Urol Nephrol       Date:  2015-04-30       Impact factor: 2.370

2.  Performance of ICU ventilators during noninvasive ventilation with large leaks in a total face mask: a bench study.

Authors:  Maria Aparecida Miyuki Nakamura; Eduardo Leite Vieira Costa; Carlos Roberto Ribeiro Carvalho; Mauro Roberto Tucci
Journal:  J Bras Pneumol       Date:  2014 May-Jun       Impact factor: 2.624

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.