Literature DB >> 8034878

Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillator.

C D Swerdlow1, T Ahern, P S Chen, C Hwang, E Gang, W Mandel, R M Kass, C T Peter.   

Abstract

OBJECTIVES: The goal of this study was to determine the incidence and clinical significance of underdetection in 125 patients treated with a tiered-therapy cardioverter-defibrillator, the Medtronic PCD.
BACKGROUND: Underdetection, distinct from undersensing, is a unique, potential complication of new algorithms that enhance specificity in tiered-therapy cardioverter-defibrillators. These algorithms may delay or prevent recognition of ventricular tachycardia even though electrograms are sensed accurately and RR intervals meet the programmed interval criterion.
METHODS: Underdetection was defined as delay in detection > 5 s at electrophysiologic study or symptomatic delay or detection failure at follow-up of 15 +/- 8 months.
RESULTS: We identified six specific mechanisms of underdetection caused by algorithms to discriminate sustained ventricular tachycardia from sinus tachycardia, atrial fibrillation, ventricular fibrillation and nonsustained ventricular tachycardia. Underdetection caused detection delays in 13 (1.9%) of 677 induced ventricular tachyarrhythmia episodes in 12 patients (9.6%). During follow-up, underdetection occurred in 7 (9.9%) of 71 patients in whom ventricular tachycardia therapies were programmed. Failure to detect ventricular tachycardia occurred in 6 (0.6%) of 988 spontaneous ventricular tachycardia episodes in four patients (5.6%); 2 episodes required external cardioversion. After defibrillator reprogramming, underdetection did not occur.
CONCLUSIONS: Algorithms to enhance specificity cause underdetection of ventricular tachycardia in a significant minority of patients with tiered-therapy cardioverter-defibrillators. Optimal programming can minimize underdetection.

Entities:  

Mesh:

Year:  1994        PMID: 8034878     DOI: 10.1016/0735-1097(94)90298-4

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  6 in total

Review 1.  Methods of minimizing inappropriate implantable cardioverter-defibrillator shocks.

Authors:  B Schaer; S Osswald
Journal:  Curr Cardiol Rep       Date:  2000-07       Impact factor: 2.931

Review 2.  Algorithms for better arrhythmia discrimination in implantable cardioverter defibrillators.

Authors:  R Schimpf; C Wolpert; B Lüderitz
Journal:  Curr Cardiol Rep       Date:  2001-11       Impact factor: 2.931

Review 3.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08

4.  [New algorithms for discrimination between supraventricular and ventricular tachyarrhythmias in patients with implantable cardioverter/defibrillator].

Authors:  J Neuzner; M Schlepper
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

5.  Postpacing Interval During Right Ventricular Overdrive Pacing to Discriminate Supraventricular from Ventricular tachycardia.

Authors:  Kivanc Yalin; Ebru Golcuk; Ekrem Bilal Karaayvaz; Tolga Aksu; Muhammet Arslane; Selma Kenar Tiryakioglu; Ahmet Kaya Bilge; Kamil Adalet
Journal:  J Atr Fibrillation       Date:  2017-08-31

6.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01
  6 in total