Preliminary experience with a hybrid nonthoracotomy defibrillating system that includes a biphasic device: comparison with a standard monophasic device using the same lead system.

A Natale1, J Sra, K Axtell, C Maglio, A Dhala, S Deshpande, M Jazayeri, A Wase, M Akhtar.   

Abstract

OBJECTIVES: This study analyzed the advantage of combining a biphasic device with a transvenous system and compared the results with those obtained with a standard monophasic device.
BACKGROUND: Available lead systems use monophasic pulses and may require lengthy intraoperative testing to achieve adequate defibrillation threshold in a conspicuous number of patients. The option of biphasic waveform may provide further benefits. However, clinical experience with a permanent implant is lacking.
METHODS: Fifty-five patients underwent testing and received a permanent implant using the Endotak lead system associated with a CPI monophasic device. The remaining 36 patients received a permanent implant with the Endotak lead system connected to a biphasic device. In both groups a subcutaneous patch was combined when needed to obtain acceptable defibrillation thresholds.
RESULTS: Biphasic pulses resulted in lower mean (+/- SD) defibrillation thresholds (monophasic 15 +/- 4.7 J vs. biphasic 12 +/- 5 J, p = 0.03) and a better implantation rate (100% biphasic vs. 89% monophasic, p = 0.07). Biphasic pulses allowed implantation with less ventricular fibrillation induction (7.4 +/- 3.2 vs. 3.5 +/- 1.8, p < 0.01) and a mean shorter procedure time (168 +/- 39 vs. 111 +/- 30 min, p < 0.01). With the biphasic waveform a greater proportion of patients met the implantation criteria with the lead system alone (83% vs. 45%, p < 0.01). When needed, the left prepectoral location of the patch electrode was always sufficient in left subscapular position was required in 15 patients in the monophasic group. Implantation of the biphasic device was associated with a shorter mean hospital stay (3.8 +/- 0.8 vs. 5.4 +/- 2.2 days, p < 0.01).
CONCLUSIONS: Incorporation of a biphasic device in a transvenous implantable cardioverter-defibrillator uniformly increases the efficacy of the system and the ease of implantation.

Entities:  

Mesh:

Year:  1994        PMID: 8034875     DOI: 10.1016/0735-1097(94)90295-x

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  6 in total

1.  A comparison of pectoral and abdominal transvenous defibrillator implantation: analysis of costs and outcomes.

Authors:  M R Gold; D Froman; N G Kavesh; R W Peters; A H Foster; S R Shorofsky
Journal:  J Interv Card Electrophysiol       Date:  1998-12       Impact factor: 1.900

2.  Optimal biphasic waveforms for internal defibrillation using a 60 muF capacitor.

Authors:  Yoshio Yamanouchi; Stéphane X Garrigue; Kent A Mowrey; Bruce L Wilkoff; Patrick J Tchou
Journal:  Exp Clin Cardiol       Date:  2002

3.  Ventricular pacing thresholds following high-energy implantable cardioverter defibrillator shocks in integrated bipolar defibrillation systems.

Authors:  Yoshio Yamanouchi; Kei Miyoshi; Sunao Kodama; Yuhei Shiga; Shunichiro Sumi; Hideya Niimura; Hideaki Toujou; Hidenori Urata
Journal:  Exp Clin Cardiol       Date:  2009

4.  A randomized comparison of defibrillation thresholds in the right ventricular outflow tract versus right ventricular apex.

Authors:  Michael Mollerus; Margaret Lipinski; Thomas Munger
Journal:  J Interv Card Electrophysiol       Date:  2008-05-10       Impact factor: 1.900

5.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01

6.  Coronary vein defibrillator coil placement in patients with high defibrillation thresholds.

Authors:  Moisés Rodríguez-Mañero; Bahij Kreidieh; Sergio H Ibarra-Cortez; Paulino Álvarez; Paul Schurmann; Amish S Dave; Miguel Valderrábano
Journal:  J Arrhythm       Date:  2018-12-03
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.