The infiltration of saliva into the multi-metallic structures on titanium implants brings different types of alloys into temporary or permanent contact. In this way a galvanic cell is established as a result of their potential difference. The galvanic cell phenomenon is compounded by another type of corrosion resulting from the geometry of the assembly: localized crevice corrosion. Fifteen galvanic couples (Ti/gold-based alloys, Ti/palladium-based alloy and Ti/non-precious alloys) were studied. Various electrochemical parameters (Ecorr, Ecommon, Ecouple corr, Ecrevice, icorr, icouple corr and Tafel slopes) were analysed. The galvanic currents measured are of the same order of magnitude (except Ti/stainless steel). They remain low. Application of the mixed-potential theory shows that titanium in coupling with the alloys studied will be under either cathodic or anodic control. According to the results obtained, an alloy that is potentially usable for superstructures in a galvanic coupling with titanium must fulfil a certain number of parameters: in a coupling, titanium must have a weak anodic polarization; the current generated by the galvanic cell must also be weak; the crevice potential must be markedly higher than the common potential.Entities: