MR detection of leakage from silicone breast implants: value of a silicone-selective pulse sequence.

D L Monticciolo1, R C Nelson, W T Dixon, J Bostwick, S Mukundan, T R Hester.   

Abstract

OBJECTIVE: The purpose of this study was to determine the value of MR imaging with a silicone-selective pulse sequence for detecting leakage from silicone breast implants. SUBJECTS AND METHODS: Women with silicone breast implants were referred for this study on the basis of clinical or imaging findings suggestive of implant rupture. Twenty-eight patients with 38 implants were examined with silicone-selective MR imaging and also underwent surgical removal of the studied implant. All but four also had mammography before MR imaging. Results of silicone-selective MR imaging for the detection of silicone leakage were compared with mammographic and surgical findings. Surgical proof was considered the gold standard.
RESULTS: Silicone-selective MR imaging showed an apparently intact implant in 21 cases; 20 of these were found to be intact at surgery. Silicone-selective MR imaging showed evidence of leakage in 17 implants, all of which showed leakage at surgery. The sensitivity for detection of leakage was 94%; the specificity was 100%. The findings of silicone-selective MR imaging and mammography were in agreement in 30 of 34 cases in which both studies were performed. In the four cases of disagreement, surgical findings agreed with MR findings in three and with mammographic findings in one. When the findings of mammography and silicone-selective MR imaging were combined, the correct status (leakage or no leakage) of all implants examined was determinable.
CONCLUSION: Silicone-selective MR imaging is highly effective for detecting leakage from silicone breast implants. Accuracy is improved when mammographic and MR findings are considered together.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8010247     DOI: 10.2214/ajr.163.1.8010247

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1.  Silicone-selective multishot echo-planar imaging for rapid MRI survey of breast implants.

Authors:  Yasuo Amano; Ritsu Aoki; Shinichiro Kumita; Tatsuo Kumazaki
Journal:  Eur Radiol       Date:  2007-01-23       Impact factor: 5.315

Review 2.  The effect of study design biases on the diagnostic accuracy of magnetic resonance imaging for detecting silicone breast implant ruptures: a meta-analysis.

Authors:  Jae W Song; Hyungjin Myra Kim; Lillian T Bellfi; Kevin C Chung
Journal:  Plast Reconstr Surg       Date:  2011-03       Impact factor: 4.730

3.  [Breast cancer. Sonography and magnetic resonance mammography].

Authors:  S Delorme
Journal:  Radiologe       Date:  2004-06       Impact factor: 0.635

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.