Accuracy assessment for three fiberoptic pulmonary artery catheters for SvO2 monitoring.

A Armaganidis1, J F Dhainaut, J L Billard, K Klouche, J P Mira, F Brunet, A T Dinh-Xuan, J Dall'Ava-Santucci.   

Abstract

OBJECTIVE: To compare values of SvO2 obtained by reflectance spectrophotometry continuous monitoring with those obtained from blood samples measurements by transmission spectrophotometry (Co-Oximetry).
DESIGN: Values of SvO2 recorded by three pulmonary artery catheters for continuous monitoring, SAT1, SAT2 and Oximetrix3 (OX3), were compared in a prospective manner to those measured on blood samples by a Co-Oximeter, using the statistical analysis of Bland and Altman.
SETTING: Adult intensive care unit in an University Hospital. PATIENTS: 37 patients admitted for acute respiratory failure and/or shock who required hemodynamic monitoring. MAIN
RESULTS: The bias (average under- or overestimation) was small for all comparative measurements: +1.3, -0.2 and +1.0 sat% for SAT1, SAT2 and OX3, respectively. However, limits of agreement were only acceptable for SAT2 (-8.3 to +7.9 sat%) and OX3 (-6.7 to +8.6 sat%), but not for SAT1 (-23.3 to +25.9 sat%). No significant drift during 24 h was found with the three catheters. However, in vitro calibration was only found acceptable for SAT2 and OX3. The results were not influenced by the numbers of wavelengths of the device (2 for SAT1 and SAT2, and 3 for OX3) nor did they correlate with any of the hemodynamic and biochemical variables tested.
CONCLUSION: For usual monitoring in the ICU, SAT2 and OX3, gave SvO2 values which are in acceptable agreement with SvO2 measured on blood samples by Co-Oximetry.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7995864     DOI: 10.1007/BF01711900

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  18 in total

1.  In vivo studies with an intravascular and intracardiac reflection oximeter.

Authors:  Y ENSON; W A BRISCOE; M L POLANYI; A COURNAND
Journal:  J Appl Physiol       Date:  1962-05       Impact factor: 3.531

2.  Perioperative evaluation of a new mixed venous oxygen saturation catheter in cardiac surgical patients.

Authors:  C G Pond; G Blessios; J Bowlin; C McCawley; D G Lappas
Journal:  J Cardiothorac Vasc Anesth       Date:  1992-06       Impact factor: 2.628

3.  Clinical evaluation of the edwards laboratories and oximetrix mixed venous oxygen saturation catheters.

Authors:  J H Karis; P D Lumb
Journal:  J Cardiothorac Anesth       Date:  1988-08

4.  Comparing methods of measurement: an alternative approach.

Authors:  K R LaMantia; T O'Connor; P G Barash
Journal:  Anesthesiology       Date:  1990-05       Impact factor: 7.892

5.  Use and interpretation of common statistical tests in method-comparison studies.

Authors:  J O Westgard; M R Hunt
Journal:  Clin Chem       Date:  1973-01       Impact factor: 8.327

6.  In vivo monitoring with a fiber optic catheter.

Authors:  J B Taylor; B Lown; M Polanyi
Journal:  JAMA       Date:  1972-08-14       Impact factor: 56.272

7.  A fiberoptic reflection oximeter.

Authors:  M L Landsman; N Knop; G Kwant; G A Mook; W G Zijlstra
Journal:  Pflugers Arch       Date:  1978-03-20       Impact factor: 3.657

8.  Three mixed venous saturation catheters in patients with circulatory shock and respiratory failure.

Authors:  J J Rouby; P Poète; L Bodin; J L Bourgeois; M Arthaud; P Viars
Journal:  Chest       Date:  1990-10       Impact factor: 9.410

9.  Leukocyte larceny: a cause of spurious hypoxemia.

Authors:  M J Fox; J S Brody; L R Weintraub
Journal:  Am J Med       Date:  1979-11       Impact factor: 4.965

10.  A comparison of two pulmonary artery mixed venous oxygen saturation catheters during the changing conditions of cardiac surgery.

Authors:  B R Hecker; D L Brown; D Wilson
Journal:  J Cardiothorac Anesth       Date:  1989-06
View more
  5 in total

Review 1.  Matching total body oxygen consumption and delivery: a crucial objective?

Authors:  Pierre Squara
Journal:  Intensive Care Med       Date:  2004-09-21       Impact factor: 17.440

Review 2.  Oxygen uptake-to-delivery relationship: a way to assess adequate flow.

Authors:  Vincent Caille; Pierre Squara
Journal:  Crit Care       Date:  2006       Impact factor: 9.097

Review 3.  Clinical relevance of data from the pulmonary artery catheter.

Authors:  Emmanuel Robin; Marion Costecalde; Gilles Lebuffe; Benoît Vallet
Journal:  Crit Care       Date:  2006       Impact factor: 9.097

Review 4.  Central venous oxygenation: when physiology explains apparent discrepancies.

Authors:  Pierre Squara
Journal:  Crit Care       Date:  2014-11-10       Impact factor: 9.097

Review 5.  The contemporary pulmonary artery catheter. Part 2: measurements, limitations, and clinical applications.

Authors:  I T Bootsma; E C Boerma; T W L Scheeren; F de Lange
Journal:  J Clin Monit Comput       Date:  2021-03-01       Impact factor: 2.502

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.