Particle image velocimetry in the investigation of flow past artificial heart valves.

W L Lim1, Y T Chew, T C Chew, H T Low.   

Abstract

Full-field measurement of instantaneous velocities in the flow field of artificial heart valves is vital as the flow is unsteady and turbulent. Particle image velocimetry (PIV) provides us the ability to do this as compared to other point measurement devices where the velocity is measured at a single point in space over time. In the development of a PIV system to investigate the flow field of artificial heart valves, many problems associated with the project arose and were subsequently resolved. Experience gained in the setting up of an environment conducive for PIV studies of artificial heart valves; from seed particle selection to refractive index matching, and the evolution of computer algorithms to satisfy the varied flow conditions in artificial heart valves are presented here. Velocity profiles and distributions are computed and drawn for a porcine tissue heart valve based on measurements with the PIV system developed.

Mesh:

Year:  1994        PMID: 7978551     DOI: 10.1007/BF02368237

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  8 in total

1.  Optimal management of a ventricular assist system. Contribution of flow visualization studies.

Authors:  J C Woodard; F D Shaffer; R D Schaub; L W Lund; H S Borovetz
Journal:  ASAIO J       Date:  1992 Jul-Sep       Impact factor: 2.872

2.  Fast tracking algorithm for multiframe particle image velocimetry data.

Authors:  B R Frieden; C K Zoltani
Journal:  Appl Opt       Date:  1989-02-15       Impact factor: 1.980

3.  The use of image processing in the investigation of artificial heart valve flow.

Authors:  K Affeld; P Walker; K Schichl
Journal:  ASAIO Trans       Date:  1989 Jul-Sep

4.  Dimensions and geometric relationships of the human aortic valve as a function of pressure.

Authors:  M Swanson; R E Clark
Journal:  Circ Res       Date:  1974-12       Impact factor: 17.367

5.  Laser anemometry measurements of steady flow past aortic valve prostheses.

Authors:  Y T Chew; H T Low; C N Lee; S S Kwa
Journal:  J Biomech Eng       Date:  1993-08       Impact factor: 2.097

6.  In-vitro fluid dynamic characteristics of the abiomed trileaflet heart valve prosthesis.

Authors:  Y R Woo; F P Williams; A P Yoganathan
Journal:  J Biomech Eng       Date:  1983-11       Impact factor: 2.097

7.  Pulsatile flow experiments on heart valve prostheses.

Authors:  K B Chandran; T L Yearwood; C J Chen; H L Falsetti
Journal:  Med Biol Eng Comput       Date:  1983-09       Impact factor: 2.602

8.  Pressure drop and velocity fields at four mechanical heart valve prostheses: Björk-Shiley Standard, Björk-Shiley Concave-Convex, Hall-Kaster and St. Jude Medical.

Authors:  K H Bruss; H Reul; J van Gilse; E Knott
Journal:  Life Support Syst       Date:  1983 Jan-Mar
  8 in total
  3 in total

1.  Application of full field optical studies for pulsatile flow in a carotid artery phantom.

Authors:  M Nemati; G B Loozen; N van der Wekken; G van de Belt; H P Urbach; N Bhattacharya; S Kenjeres
Journal:  Biomed Opt Express       Date:  2015-09-21       Impact factor: 3.732

2.  Blood damage through a bileaflet mechanical heart valve: a quantitative computational study using a multiscale suspension flow solver.

Authors:  B Min Yun; Cyrus K Aidun; Ajit P Yoganathan
Journal:  J Biomech Eng       Date:  2014-10       Impact factor: 2.097

3.  Transcatheter aortic valves produce unphysiological flows which may contribute to thromboembolic events: An in-vitro study.

Authors:  Andrea Ducci; Francesco Pirisi; Spyridon Tzamtzis; Gaetano Burriesci
Journal:  J Biomech       Date:  2016-11-03       Impact factor: 2.712

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.