Literature DB >> 7960623

Accuracy of volume measurements of computed tomography and magnetic resonance imaging phantoms by three-dimensional reconstruction and preliminary clinical application.

D G Disler1, D S Marr, D I Rosenthal.   

Abstract

RATIONALE AND
OBJECTIVES: Using an independent three-dimensional workstation, the accuracy of volume measurements of phantoms was assessed using three-dimensional reconstruction of two-dimensional computed tomography (CT) and magnetic resonance (MR) images.
METHODS: Round, cylindrical, and irregularly shaped high-contrast phantoms of known volume were imaged in a water bath. The effect of object contrast on volume estimation was tested using phantoms of known volume diluted serially with a contrast agent. The effect of changing field of view and slice thickness was assessed. A clinical application was performed, in which nine shoulders were injected with a known quantity of contrast material, to test the accuracy of the technique in vivo.
RESULTS: A strong paired correlation (r = .99) between estimated and true volumes was obtained for high-contrast phantoms ranging from 17 to 128 mL. The weighted average absolute error was 1.42 mL (MR) and 3.50 mL (CT). Accuracy of the serially diluted 27-mL phantoms was essentially unaffected by contrast differences greater than 133 units (MR) and 102 units (CT). The weighted average absolute error was 1.33 mL (MR) and 1.56 mL (CT). Changing field of view had no effect on accuracy, but increasing the slice thickness resulted in over-estimation of volume. The mean error for the clinical application was 4.4% (range: 1.7%-8.3%).
CONCLUSION: Under certain circumstances, three-dimensional reconstructive volume estimation can be a convenient and accurate method for volume determination.

Entities:  

Mesh:

Year:  1994        PMID: 7960623     DOI: 10.1097/00004424-199408000-00002

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  10 in total

1.  Renal volume assessment with 3D ultrasound.

Authors:  A Brancaforte; S Serantoni; F Silva Barbosa; G Di Leo; F Sardanelli; G P Cornalba
Journal:  Radiol Med       Date:  2011-06-04       Impact factor: 3.469

2.  Reliability of CT-based tumor volumetry after intraarterial chemotherapy in patients with small carcinoma of the oral cavity and the oropharynx.

Authors:  Stefan Rohde; Adorján F Kovács; Joachim Berkefeld; Bernd Turowski
Journal:  Neuroradiology       Date:  2006-04-12       Impact factor: 2.804

3.  Validation of multi-detector computed tomography as a non-invasive method for measuring ovarian volume in macaques (Macaca fascicularis).

Authors:  Jeryl C Jones; Susan E Appt; Stephen R Werre; Joshua C Tan; Jay R Kaplan
Journal:  Am J Primatol       Date:  2010-06       Impact factor: 2.371

4.  The change in ratio of convex and concave lung volume in adolescent idiopathic scoliosis: a 3D CT scan based cross sectional study of effect of severity of curve on convex and concave lung volumes in 99 cases.

Authors:  Eun Mi Chun; Seung Woo Suh; Hitesh N Modi; Eun Young Kang; Suk Joo Hong; Hae-Ryong Song
Journal:  Eur Spine J       Date:  2007-10-03       Impact factor: 3.134

5.  Subchondral bone and cartilage thickness from MRI: effects of chemical-shift artifact.

Authors:  Chris A McGibbon; Jenny Bencardino; William E Palmer
Journal:  MAGMA       Date:  2003-02       Impact factor: 2.310

6.  From 2D to 3D: an algorithm to derive normal values for 3-dimensional computerized assessment.

Authors:  Bruno Frazäo Gribel; Marcos Nadler Gribel; Flavio Ricardo Manzi; Sharon L Brooks; James A McNamara
Journal:  Angle Orthod       Date:  2011-01       Impact factor: 2.079

7.  Preclinical multimodality phantom design for quality assurance of tumor size measurement.

Authors:  Yongsook C Lee; Gary D Fullerton; Cristel Baiu; Margaret G Lescrenier; Beth A Goins
Journal:  BMC Med Phys       Date:  2011-09-30

8.  Pancreas volumes in humans from birth to age one hundred taking into account sex, obesity, and presence of type-2 diabetes.

Authors:  Y Saisho; A E Butler; J J Meier; T Monchamp; M Allen-Auerbach; R A Rizza; P C Butler
Journal:  Clin Anat       Date:  2007-11       Impact factor: 2.414

9.  MR evaluation of radiation synovectomy of the knee by means of intra-articular injection of holmium-166-chitosan complex in patients with rheumatoid arthritis: results at 4-month follow-up.

Authors:  Sang Hoon Lee; Jin Suck Suh; Ho Seok Kim; Jong Doo Lee; Jungsik Song; Soo Kon Lee
Journal:  Korean J Radiol       Date:  2003 Jul-Sep       Impact factor: 3.500

10.  Volumetric accuracy of cone-beam computed tomography.

Authors:  Cheol-Woo Park; Jin-Ho Kim; Yu-Kyeong Seo; Sae-Rom Lee; Ju-Hee Kang; Song-Hee Oh; Gyu-Tae Kim; Yong-Suk Choi; Eui-Hwan Hwang
Journal:  Imaging Sci Dent       Date:  2017-09-21
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.