Extrinsic factors influencing left ventricular conductance in situ.

M M Amirhamzeh1, C X Jia, H M Spotnitz.   

Abstract

BACKGROUND: The accuracy of conductance measurement of left ventricular (LV) volume is affected by parallel conductance (Vp) from right ventricular (RV) volume as well as surroundings of the heart. No studies have quantified the extrinsic influences on Vp. METHODS AND
RESULTS: In six anesthetized pigs, using a median sternotomy and pericardial well, implantable cardioverter defibrillator patches, normal saline, or echocardiography gel were introduced and removed. LV end-diastolic pressure (LVEDP) did not change significantly with these interventions. Raw LV conductance was used to estimate end-diastolic volume (LVEDV), end-systolic volume (LVESV), and stroke volume (SV). Saline significantly increased mean +/- SEM calculated LVEDV from 106 +/- 8 to 131 +/- 6 mL and calculated LVESV from 81 +/- 7 to 110 +/- 5 mL; calculated SV decreased from 25 +/- 3 to 21 +/- 2 mL (P < .05, paired t test). Implantable cardioverter defibrillator patches and gel did not significantly change calculated LV volumes. LV conductance was also measured in arrested hearts during LV filling (5 to 60 mL) at four levels of RV volume (RVV). Using a fixed value for Vp, calculated mean LV volume was 30 +/- 3 mL at RVV = 0, 32 +/- 3 mL at RVV = 20, 33 +/- 3 mL at RVV = 40, and 36 +/- 3 mL at RVV = 60 (P < .01, repeated-measures ANOVA).
CONCLUSIONS: While implantable cardioverter defibrillator patches and gel have no statistically significant effect, saline and RVV affect LV volume calculations by conductance, requiring special techniques during cardiopulmonary bypass or caval occlusion.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7955277

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  2 in total

1.  Parallel conductance determination in cardiac volumetry using dilution manoeuvres: theoretical analysis and practical implications.

Authors:  M C Herrera; J M Olivera; M E Valentinuzzi
Journal:  Med Biol Eng Comput       Date:  1999-03       Impact factor: 2.602

2.  In vitro and finite-element model investigation of the conductance technique for measurement of aortic segmental volume.

Authors:  D A Hettrick; J H Battocletti; J J Ackmann; J H Linehan; D C Warltier
Journal:  Ann Biomed Eng       Date:  1996 Nov-Dec       Impact factor: 3.934

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.