A study of the relationship between patient dose and size in paediatric radiology.

C J Martin1, B Farquhar, E Stockdale, S MacDonald.   

Abstract

Entrance doses and dose-area products have been measured for paediatric X-ray examinations in a dedicated paediatric department. Dose-area products for barium meal and micturating cystourethrogram examinations involving fluoroscopy were substantially less than results from other studies. A number of factors contribute to this, including a high gain image intensifier used without a grid, shorter screening times, smaller numbers of films, exclusive use of a 100 mm camera and careful collimation of the X-ray field. Entrance air kerma has been measured for pelvis, abdomen, chest and skull radiographs. The relationship between dosimetric quantities and an equivalent patient diameter derived from data on patient height and weight has been studied. The results show a linear relationship between the logarithm of entrance air kerma and equivalent patient diameter for radiographic examinations performed on a single X-ray unit. Application of exponential correction factors could allow entrance doses for radiographs of children of different ages to be compared with appropriate reference doses.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7953228     DOI: 10.1259/0007-1285-67-801-864

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  8 in total

1.  Reference levels for patient radiation doses in interventional radiology: proposed initial values for U.S. practice.

Authors:  Donald L Miller; Deukwoo Kwon; Grant H Bonavia
Journal:  Radiology       Date:  2009-09-29       Impact factor: 11.105

2.  Impact of StentBoost subtract imaging on patient radiation exposure during percutaneous coronary intervention.

Authors:  Zhigeng Jin; Shengli Yang; Limin Jing; Huiliang Liu
Journal:  Int J Cardiovasc Imaging       Date:  2013-03-03       Impact factor: 2.357

3.  Radiation exposure and patient experience during percutaneous coronary intervention using radial and femoral artery access.

Authors:  Håkan Geijer; Jan Persliden
Journal:  Eur Radiol       Date:  2004-04-21       Impact factor: 5.315

4.  Radiation dose and cancer risk to children undergoing skull radiography.

Authors:  Michael Mazonakis; John Damilakis; Maria Raissaki; Nicholas Gourtsoyiannis
Journal:  Pediatr Radiol       Date:  2004-06-12

5.  A method to derive appropriate exposure parameters from target exposure index and patient thickness in pediatric digital radiography.

Authors:  Menglong Zhang; Kai Liu; Xuecai Niu; Xinli Liu
Journal:  Pediatr Radiol       Date:  2012-12-04

6.  Evaluation of radiation dose and image quality following changes to tube potential (kVp) in conventional paediatric chest radiography.

Authors:  S Ramanaidu; Rb Sta Maria; Kh Ng; J George; G Kumar
Journal:  Biomed Imaging Interv J       Date:  2006-07-01

7.  Gonad shielding in paediatric pelvic radiography: disadvantages prevail over benefit.

Authors:  Marij J Frantzen; Simon Robben; Alida A Postma; Johannes Zoetelief; Joachim E Wildberger; Gerrit J Kemerink
Journal:  Insights Imaging       Date:  2011-09-25

8.  Foetal radiation dose and risk from diagnostic radiology procedures: a multinational study.

Authors:  Ernest K Osei; Johnson Darko
Journal:  ISRN Radiol       Date:  2012-09-25
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.