A practical and reliable method of measuring blood pressure in the neonate by pulse oximetry.

M Langbaum1, F G Eyal.   

Abstract

OBJECTIVE: We investigated the reliability of the plethysmographic waveform of the pulse oximeter to measure systolic blood pressure in sick neonates.
METHODS: Fifty infants admitted to the neonatal intensive care unit, with indwelling arterial catheters placed for their ongoing care, were enrolled. Median gestational age was 31 weeks (range, 24 to 40 weeks), and the mean birth weight was 1711 gm (range, 546 to 3856 gm). Blood pressure was recorded by an oscillometric method as well as from a transducer connected to an arterial catheter. Additionally, pulse oximeter blood pressures were obtained by gradually inflating an appropriately sized blood pressure cuff in increments of 2 to 5 mm Hg, on the same extremity as the oximeter probe, until the waveform just disappeared. The cuff was then rapidly inflated another 20 mm Hg and then gradually deflated in increments of 2 to 5 mm Hg until the waveform reappeared on the oximeter screen display. The pulse oximeter blood pressures were calculated both as the blood pressure noted at disappearance of the pulse oximeter waveform and as the blood pressure noted by the average pulse oximeter blood pressure at the disappearance and reappearance of the waveform. The mean intraarterial systolic blood pressure was 54 mm Hg (range, 36 to 82 mm Hg).
RESULTS: Blood pressures obtained by pulse oximetry showed a significantly better correlation with intraarterial measurements in comparison with those obtained by oscillometric instruments. Additionally, the limits of agreement (mean difference +/- 2 SD) between blood pressures obtained by intraarterial measurements and those obtained by pulse oximetry were within a clinically acceptable range as opposed to those obtained by the comparison of intraarterial and oscillometric methods.
CONCLUSION: Measurements of blood pressure in the neonate by means of pulse oximetry waveform analysis are easily obtainable and more accurate than those obtained by the oscillometric method.

Entities:  

Mesh:

Year:  1994        PMID: 7931880     DOI: 10.1016/s0022-3476(94)70016-8

Source DB:  PubMed          Journal:  J Pediatr        ISSN: 0022-3476            Impact factor:   4.406


  9 in total

Review 1.  Hypertension in the newborn baby.

Authors:  M Watkinson
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-03       Impact factor: 5.747

2.  Non-invasive measurement of systolic blood pressure on the arm utilising photoplethysmography: development of the methodology.

Authors:  C Laurent; B Jönsson; M Vegfors; L G Lindberg
Journal:  Med Biol Eng Comput       Date:  2005-01       Impact factor: 2.602

Review 3.  Monitoring during paediatric cardiac anaesthesia.

Authors:  J P Purday
Journal:  Can J Anaesth       Date:  1994-09       Impact factor: 5.063

4.  Systolic blood pressure in babies of less than 32 weeks gestation in the first year of life. Northern Neonatal Nursing Initiative.

Authors: 
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-01       Impact factor: 5.747

5.  Automatic noninvasive measurement of systolic blood pressure using photoplethysmography.

Authors:  Meir Nitzan; Amikam Patron; Zehava Glik; Abraham T Weiss
Journal:  Biomed Eng Online       Date:  2009-10-26       Impact factor: 2.819

6.  Use of pulse oximetry for blood pressure measurement after cardiac surgery.

Authors:  A J Movius; S L Bratton; G K Sorensen
Journal:  Arch Dis Child       Date:  1998-05       Impact factor: 3.791

7.  Pulse oximetry vs non-invasive blood pressure/oscillometry to record blood pressure in neonates: A prospective observational study.

Authors:  Reenu Raju; Mathai Isac
Journal:  J Family Med Prim Care       Date:  2022-06-30

8.  Comparison of systolic blood pressure values obtained by photoplethysmography and by Korotkoff sounds.

Authors:  Meir Nitzan; Yair Adar; Ellie Hoffman; Eran Shalom; Shlomo Engelberg; Iddo Z Ben-Dov; Michael Bursztyn
Journal:  Sensors (Basel)       Date:  2013-10-31       Impact factor: 3.576

9.  More accurate systolic blood pressure measurement is required for improved hypertension management: a perspective.

Authors:  Meir Nitzan; Itzchak Slotki; Linda Shavit
Journal:  Med Devices (Auckl)       Date:  2017-07-24
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.