| Literature DB >> 7927390 |
Y C Okada1, B Shah, J C Huang.
Abstract
A low cost, two-shell ferromagnetic shielded room large enough for a person to enter and prepare experiments was constructed for biomagnetic measurements. No aluminium or copper shell was used for eddy-current shielding. A high-permeability nickel-iron-molybdenum alloy (1.57 mm thick) was used for both ferromagnetic and eddy-current shieldings. The shielding factor was 60 dB at dc and 40 dB between 0.1 Hz and about 3 Hz. The eddy-current shielding due to the alloy alone provided a shielding factor of 55 dB at 30 Hz and 66 dB at 60 Hz. The shielding factor was sufficiently high in all the frequency range for biomagnetic measurements with a first-order superconducting gradiometer.Entities:
Mesh:
Substances:
Year: 1994 PMID: 7927390 DOI: 10.1109/10.301736
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538