Flow pattern analysis of the Baylor total artificial heart.

D Meier1, J T Wernicke, Y Orime, S Takatani, K Tasai, G Damm, K Naito, K Mizuguchi, K Makinouchi, J Glueck.   

Abstract

To obtain an optimal design of the left blood chamber of the total artificial heart (TAH), flow visualization studies were performed. Information on velocities in critical areas of the left chamber was gathered using sectional pulsed laser light. The flow patterns on the entire pumping duration were photographed frame by frame. The inflow port, the opposite of the inflow and outflow of the pump (bottom area), and the diaphragm/housing junction were the focal areas. The pump conditions, such as chamber pressure, preload and afterload pressure, pumping rate and roller screw, and displacement were recorded. Major stagnations and a low washout effect were observed in the bottom region. The closing of the inflow valve was irregular. In order to solve this problem, the inflow valve angle was changed 20 degrees. A comparison study showed a better valve closing characteristic, and no stagnation areas were observed with this new valve angle. Various velocity profiles confirmed the results. The valve closing characteristics is described in relationship to back flow.

Mesh:

Year:  1994        PMID: 7887830     DOI: 10.1111/j.1525-1594.1994.tb03345.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  1 in total

1.  Flow visualization for different port angles of a pulsatile ventricular assist device.

Authors:  Eiki Akagawa; Hwansung Lee; Eisuke Tatsumi; Akihiko Homma; Tomonori Tsukiya; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2011-10-30       Impact factor: 1.731

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.