Volume detection threshold: quantitative comparison of computed radiography and screen-film radiography in detection of pneumothoraces in an animal model that simulates the neonate.

S Don1, M D Cohen, R A Kruger, T A Winkler, B P Katz, W Li, R G Dreesen, N Kennan, R Tarver, E C Klatte.   

Abstract

PURPOSE: To quantitatively compare computed radiography (CR) and screen-film radiography (SFR) in the detection of pneumothorax in an animal model that simulates the neonate.
MATERIALS AND METHODS: Three rabbits underwent placement of 5-F catheters in the right pleural space. Eight CR and eight SFR images were obtained. Volume-controlled pneumothorax was induced by incrementally injecting 4 cm3 of air (24 cm3 total) and four CR and four SFR images were obtained after each injection. Four radiologists blindly viewed the images twice (eight readings per image, 1,600 total observations). A dichotomous yes-no score for the presence of pneumothorax was used to calculate the volume detection threshold (VDT), defined as the air volume at which 50% of the pneumothoraces can be detected. The authors plotted the likelihood of a pneumothorax against the air volume by using the confidence indicator.
RESULTS: Intraobserver VDT values for CR and SFR were not significantly different.
CONCLUSION: CR is as accurate as SFR in the detection of pneumothorax in this quantitative animal model.

Entities:  

Mesh:

Year:  1995        PMID: 7862970     DOI: 10.1148/radiology.194.3.7862970

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  1 in total

1.  Computed radiography in neonatal and pediatric intensive care units: a comparison of 2.5 K x 2 K soft-copy images vs digital hard-copy film.

Authors:  P W Brill; P Winchester; P Cahill; M Lesser; S M Durfee; C S Giess; P A Auld; B Greenwald
Journal:  Pediatr Radiol       Date:  1996
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.