Infrared tympanic thermometry for neonatal temperature assessment.

M E Weiss1, D Poeltler, I Gocka.   

Abstract

OBJECTIVE: To investigate the accuracy and precision of infrared tympanic thermometer use with neonates by comparing with axillary and rectal measurements.
DESIGN: Descriptive, comparative study.
SETTING: Newborn nursery of a tertiary-level perinatal center.
SUBJECTS: Thirty-four full-term newborns. INSTRUMENTS: Infrared thermometers were used to collect tympanic temperatures. An electronic thermometer was used to collect axillary and rectal temperatures.
RESULTS: There was no significant difference between tympanic and axillary temperatures. The range of variation (95% CI) was +/- 0.7-.9 degrees C. The mean difference between tympanic and rectal measurements was 0.4 degrees C. The range of variation was +/- 0.7-0.75 degrees C. Right ear measurements produced the best approximation of axillary temperature, and protected ear measurements produced the best approximation of rectal temperature. The protected ear was 0.2-0.3 degrees C higher than the exposed ear.
CONCLUSIONS: There was more variation between tympanic versus axillary and tympanic versus rectal measurements than between axillary and rectal measurements. Tympanic thermometry may be useful for rapid screening of neonatal temperature, but its usefulness for monitoring unstable neonates remains in question.

Mesh:

Year:  1994        PMID: 7853086     DOI: 10.1111/j.1552-6909.1994.tb01955.x

Source DB:  PubMed          Journal:  J Obstet Gynecol Neonatal Nurs        ISSN: 0090-0311


  3 in total

1.  Temporal Artery Temperature Measurement in the Neonate.

Authors:  Mashette E Syrkin-Nikolau; Karen J Johnson; Tarah T Colaizy; Ruthann Schrock; Edward F Bell
Journal:  Am J Perinatol       Date:  2017-04-10       Impact factor: 1.862

2.  A comparison of tympanic and rectal temperatures in term Nigerian neonates.

Authors:  Chika O Duru; Felix O Akinbami; Adebola E Orimadegun
Journal:  BMC Pediatr       Date:  2012-06-25       Impact factor: 2.125

3.  Intelligent neonatal monitoring based on a virtual thermal sensor.

Authors:  Abbas K Abbas; Steffen Leonhardt
Journal:  BMC Med Imaging       Date:  2014-03-02       Impact factor: 1.930

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.