Phased array RF coils for high-resolution MRI of the inner ear and brain stem.

P Schmalbrock1, J Pruski, L Sun, A Rao, J W Monroe.   

Abstract

OBJECTIVE: The spatial resolution in MRI is predominantly limited by the available signal-to-noise ratio (S/N). To increase the S/N, a four coil phased array consisting of bilateral pairs of semioval coils was constructed for high resolution imaging of the temporal bone and brainstem.
MATERIALS AND METHODS: Coil sizes of 10 x 6 and 6 x 4.5 cm were tested. The S/N values were measured in vivo and with homogeneous phantoms and compared to commercial 3 in receive and quadrature head coils.
RESULTS: At a depth of 4-5 cm, phantom studies yielded S/N improvements of a factor of 1.26-1.37 with the large array compared to the 3 in coil and 2.33-1.74 compared to the head coil. Similar improvements (1.16 and 2.37) were obtained in inner ear images. No further improvement was achieved at this depth with the small array. At a depth of 8 cm, phantom studies yielded similar S/N for the quadrature head coil and two bilaterally placed large array coils, while a factor of 1.27 was obtained in brainstem images.
CONCLUSION: The use of phased array coils yields significantly increased S/N and is thus valuable for high resolution MRI.

Mesh:

Year:  1995        PMID: 7822553     DOI: 10.1097/00004728-199501000-00002

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  1 in total

1.  Signal-to-noise ratio and parallel imaging performance of commercially available phased array coils in 3.0 T brain magnetic resonance imaging.

Authors:  Tsukasa Yoshida; Kensei Shirata; Atsushi Urikura; Michitoshi Ito; Yoshihiro Nakaya
Journal:  Radiol Phys Technol       Date:  2015-05-26
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.