Comparison of sensititre broth microdilution and agar dilution susceptibility testing techniques for meropenem to determine accuracy, reproducibility, and predictive values.

M J Dowzicky1, H L Nadler, W Sheikh.   

Abstract

The stability, accuracy, reproducibility, and predictive value of Sensititre MIC panels containing meropenem (Merrem) were evaluated by using National Committee for Clinical Laboratory Standards (NCCLS)-recommended American Type Culture Collection (ATCC) strains and 110 selected strains of rapidly growing and fastidious aerobes and anaerobes with various degrees of susceptibility to meropenem. The NCCLS-recommended agar dilution method was used as a standard reference method. Meropenem-containing Sensititre MIC panels were monitored for their stabilities at room temperature and reproducibilities over 24 months by using six ATCC strains. Ninety-nine percent of the MICs of both meropenem and imipenem obtained for NCCLS-recommended ATCC strains were within the established ranges after 2 years. The overall agreement (+/- 1 twofold dilution) between the Sensititre and the agar dilution meropenem MICs was greater than 93%. The predictive value of meropenem MICs for indicating suspeptibility or resistance obtained by the Sensititre method was greater than 90%. No major or very major interpretive errors were observed, and only 5% of meropenem MICs were associated with minor interpretive errors. Problematic organisms were not observed. The Sensititre MIC panels containing meropenem offer a convenient and valid alternative to the NCCLS reference method for the susceptibility testing of potential pathogens likely to be recovered from mixed infections.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7814547      PMCID: PMC263967          DOI: 10.1128/jcm.32.9.2204-2207.1994

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  5 in total

1.  External factors affecting imipenem performance in dried microdilution MIC plates.

Authors:  R Grist
Journal:  J Clin Microbiol       Date:  1992-02       Impact factor: 5.948

2.  False resistance to imipenem with a microdilution susceptibility testing system.

Authors:  E J O'Rourke; K G Lambert; K C Parsonnet; A B Macone; D A Goldmann
Journal:  J Clin Microbiol       Date:  1991-04       Impact factor: 5.948

3.  Quality control criteria for testing the susceptibility of anaerobic bacteria to meropenem.

Authors:  R J Zabransky; L Bayola-Mueller; S G Jenkins; D R Pitkin; P C Schreckenberger; D M Celig; C A Spiegel
Journal:  J Clin Microbiol       Date:  1990-12       Impact factor: 5.948

4.  Effects of storage temperature and pH on the stability of eleven beta-lactam antibiotics in MIC trays.

Authors:  D J Nickolai; C J Lammel; B A Byford; J H Morris; E B Kaplan; W K Hadley; G F Brooks
Journal:  J Clin Microbiol       Date:  1985-03       Impact factor: 5.948

5.  Pseudoresistance of Pseudomonas aeruginosa resulting from degradation of imipenem in an automated susceptibility testing system with predried panels.

Authors:  R L White; M B Kays; L V Friedrich; E W Brown; J R Koonce
Journal:  J Clin Microbiol       Date:  1991-02       Impact factor: 5.948

  5 in total
  2 in total

1.  Validation of the automated reading and incubation system with Sensititre plates for antimicrobial susceptibility testing.

Authors:  Kimberle C Chapin; Michael C Musgnug
Journal:  J Clin Microbiol       Date:  2003-05       Impact factor: 5.948

2.  In vitro Degradation of Antimicrobials during Use of Broth Microdilution Method Can Increase the Measured Minimal Inhibitory and Minimal Bactericidal Concentrations.

Authors:  Elodie A Lallemand; Marlène Z Lacroix; Pierre-Louis Toutain; Séverine Boullier; Aude A Ferran; Alain Bousquet-Melou
Journal:  Front Microbiol       Date:  2016-12-21       Impact factor: 5.640

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.