B S Trembly1, E B Douple, T P Ryan, P J Hoopes.
Abstract
Perfused, canine skeletal muscle and the brain tumour of a cancer patient were heated with an array of four parallel, interstitial antennas placed on the corners of a 2-cm square and driven at 915 MHz. The temperature distributions along the axial and diagonal catheters were measured with equal-phase driving of the antennas and with several time-varying schemes of driving phase differences among the antennas. When equal-phase driving was replaced by a rotating scheme of 90 degrees driving phase differences, the tissue area in the junction plane heated above a normalized index temperature of 0.6 increased by a factor of about 1.25. With a rotating phase of 135 degrees, the same area increased by a factor of about 1.6. The axial temperature distribution was not affected significantly by driving phase.Entities:
Mesh:
Year: 1994 PMID: 7806925 DOI: 10.3109/02656739409022448
Source DB: PubMed Journal: Int J Hyperthermia ISSN: 0265-6736 Impact factor: 3.914