BACKGROUND: Computerised x-ray planimetry has been advocated as an alternative to body plethysmography and helium dilution for measuring static lung volumes. The accuracy and reproducibility of this method has been assessed in comparison with these standard methods.
METHODS: Plethysmographic and planimetric measurements of total lung capacity (TLC) and functional residual capacity (FRC) were made in 10 normal subjects and in 12 patients with chronic obstructive pulmonary disease (COPD), with additional helium dilution measurements in the latter 12 patients.
RESULTS: Mean lung volumes (TLC and FRC) for groups of subjects measured by planimetry and by plethysmography were similar in both groups and larger than the helium dilution measurement in patients with COPD. Intraindividual agreement between planimetry and plethysmography was poor, however, with a wide confidence interval (-2.2 to +2.31). The planimeter did not measure reliably changes in volume from TLC to FRC in individuals.
CONCLUSIONS: Mean lung volumes measured by planimetry in a group of patients probably reflect a regression to the mean of the computer algorithm rather than accurate TLC estimation. The technique is not yet robust enough to replace the established techniques of helium dilution or plethysmography.

Entities:  

Mesh:

Year:  1995        PMID: 7785011      PMCID: PMC474286          DOI: 10.1136/thx.50.4.383

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  14 in total

1.  A rapid plethysmographic method for measuring thoracic gas volume: a comparison with a nitrogen washout method for measuring functional residual capacity in normal subjects.

Authors:  A B DUBOIS; S Y BOTELHO; G N BEDELL; R MARSHALL; J H COMROE
Journal:  J Clin Invest       Date:  1956-03       Impact factor: 14.808

2.  STUDIES OF TOTAL PULMONARY CAPACITY AND ITS SUB-DIVISIONS. II. CORRELATION WITH PHYSICAL AND RADIOLOGICAL MEASUREMENTS.

Authors:  A Hurtado; W W Fray
Journal:  J Clin Invest       Date:  1933-09       Impact factor: 14.808

3.  Determination of lung volumes from chest films.

Authors:  J O'Shea; N L Lapp; A D Russakoff; R Reger; W K Morgan
Journal:  Thorax       Date:  1970-09       Impact factor: 9.139

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Frequency dependence of plethysmographic volume in healthy and asthmatic subjects.

Authors:  D O Rodenstein; D C Stănescu
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1983-01

6.  Radiographic and plethysmographic determination of total lung capacity.

Authors:  H M Loyd; S T String; A B DuBois
Journal:  Radiology       Date:  1966-01       Impact factor: 11.105

7.  A method for the determination of total lung capacity from posteroanterior and lateral chest roentgenograms.

Authors:  P C Pratt; G A Klugh
Journal:  Am Rev Respir Dis       Date:  1967-09

8.  Failure of body plethysmography in bronchial asthma.

Authors:  D C Stănescu; D Rodenstein; M Cauberghs; K P Van de Woestijne
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1982-04

9.  Predictive equations for total lung capacity and residual volume calculated from radiographs in a random sample of the Michigan population.

Authors:  K H Kilburn; R H Warshaw; J C Thornton; K Thornton; A Miller
Journal:  Thorax       Date:  1992-07       Impact factor: 9.139

10.  Reassessment of lung volume measurement by helium dilution and by body plethysmography in chronic air-flow obstruction.

Authors:  D O Rodenstein; D C Stănescu
Journal:  Am Rev Respir Dis       Date:  1982-12
View more
  2 in total

1.  Alternative methods for assessing bronchodilator reversibility in chronic obstructive pulmonary disease.

Authors:  J Hadcroft; P M Calverley
Journal:  Thorax       Date:  2001-09       Impact factor: 9.139

2.  Can the single-breath helium dilution method predict lung volumes as measured by whole-body plethysmography?

Authors:  Patrícia Chaves Coertjens; Marli Maria Knorst; Anelise Dumke; Adriane Schmidt Pasqualoto; João Riboldi; Sérgio Saldanha Menna Barreto
Journal:  J Bras Pneumol       Date:  2013 Nov-Dec       Impact factor: 2.624

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.