[Evaluation of 3 automatic systems for measurement of the erythrocyte sedimentation rate].

I Besson1, M Kinder, J M Jou, J L Vives Corrons.   

Abstract

PURPOSE: To evaluate three automated devices for measuring the erythrocyte sedimentation rate (ESR) (VES-MATIC 60, Menarini(R); SEDISCAN Becton-Dickinson(R) y SEDIMATIC, Ral(R)) by comparison with the Westergren method (WM).
MATERIAL AND METHODS: A total of 1576 whole blood samples (VM: 694, SC: 316 and SM: 566) from patients of the Hospital Clínic i Provincial de Barcelona were included in this study. In all the specimens, the ESR was determined following the ICSH recommendations (WM). The Student's t test for paired data and linear regression analysis were used for inaccuracy study. Reproducibility was assessed after five measurements of three different samples and establishing the coefficient of variation (CV).
RESULTS: Significant correlation was found between the systems studied and th WM. Moreover, for ESR > 21 mm/h (groups II, III and IV) the results provided by VM system were not significantly different from those of WM. Finally, all the systems presented a good reproducibility, although the lower values of CV were obtained with the VM method.
CONCLUSIONS: The automatic systems for measurements of the ESR demonstrate important advantages and, from this analysis, we concluded that the VM could be the alternative method to conventional Westergren for the ESR determination.

Entities:  

Mesh:

Year:  1995        PMID: 7784938

Source DB:  PubMed          Journal:  Sangre (Barc)        ISSN: 0036-4355


  2 in total

1.  Three-way comparison of methods for the measurement of the erythrocyte sedimentation rate.

Authors:  Johnny Ndoni Mahlangu; Melony Davids
Journal:  J Clin Lab Anal       Date:  2008       Impact factor: 2.352

2.  Differences in erythrocyte sedimentation rates using a modified Westergren method and an alternate method.

Authors:  Elise Schapkaitz; Shilla RabuRabu; Marcel Engelbrecht
Journal:  J Clin Lab Anal       Date:  2018-08-26       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.