Comparison of pressure and flow triggering systems during continuous positive airway pressure.

R D Branson1, R S Campbell, K Davis, D J Johnson.   

Abstract

STUDY
OBJECTIVE: Compare the inspiratory work of breathing (WOBI) during pressure triggering (PT), and flow triggering (FT) using two microprocessor ventilators.
DESIGN: A randomized, crossover comparison of triggering strategies and ventilators used.
SETTING: Surgical intensive care unit. PATIENTS: Ten patients recovering from acute respiratory failure (eight men, two women; mean age, 48 +/- 12 years) breathing on continuous positive airway pressure (CPAP) of 5 cm H2O were studied.
INTERVENTIONS: Patients were randomly assigned to either receive 5 cm H2O CPAP via one of two units (Hamilton Veolar or Puritan Bennett 7200ae) using either PT or FT. Each patient had 30-min trials using the following: (1) Veolar FT; (2) Veolar PT; (3) 7200ae FT; and (4) 7200ae FT. MEASUREMENTS AND
RESULTS: During each trial period, work of breathing (WOB) and pressure time product (PTP) were measured using a respiratory monitor (Bi-core CP-100). All patients were placed in semi-Fowler position and esophageal balloons were inserted and their position confirmed using the occlusion technique. Continuous measurements of peak negative pressure during inspiration, tidal volume (VT), minute ventilation (VE), respiratory frequency (f) were accomplished with a flow transducer at the proximal airway. FT with the 7200ae was superior to PT as measured by both the WOB (0.58 +/- 0.3 vs 0.84 +/- 0.2 J/L, p < 0.01) and PTP (148 +/- 50 vs 206 +/- 41 cm H2O/s/min, p, 0.05). FT with the Veolar was also superior to PT with respect to the WOB (0.53 +/- 0.2 vs 0.93 +/- 0.2 J/L, p < 0.01) and PTP (140 +/- 39 vs 229 +/- 46 cm H2O/s/min, p < 0.05).
CONCLUSION: FT reduces the WOB compared with PT, regardless of the ventilator used. The reduction in WOB during FT is related to improved responsiveness and changes in the posttrigger phase.

Entities:  

Mesh:

Year:  1994        PMID: 7774334     DOI: 10.1378/chest.106.2.540

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  6 in total

1.  A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators.

Authors:  Arnaud W Thille; Aissam Lyazidi; Jean-Christophe M Richard; Fabrice Galia; Laurent Brochard
Journal:  Intensive Care Med       Date:  2009-04-08       Impact factor: 17.440

Review 2.  Techniques in mechanical ventilation: principles and practice.

Authors:  J M Shneerson
Journal:  Thorax       Date:  1996-07       Impact factor: 9.139

Review 3.  Bedside waveforms interpretation as a tool to identify patient-ventilator asynchronies.

Authors:  Dimitris Georgopoulos; George Prinianakis; Eumorfia Kondili
Journal:  Intensive Care Med       Date:  2005-11-09       Impact factor: 17.440

4.  Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support.

Authors:  George Prinianakis; Eumorfia Kondili; Dimitris Georgopoulos
Journal:  Intensive Care Med       Date:  2003-03-29       Impact factor: 17.440

5.  Work of breathing, inspiratory flow response, and expiratory resistance during continuous positive airway pressure with the ventilators EVITA-2, EVITA-4 and SV 300.

Authors:  E Calzia; K H Lindner; W Stahl; A Martin; P Radermacher; M Georgieff
Journal:  Intensive Care Med       Date:  1998-09       Impact factor: 17.440

6.  Application of neurally adjusted ventilatory assist in ventilator weaning of infants ventilator weaning.

Authors:  Shuna Xiao; Chengjiao Huang; Ying Cheng; Zhi Xia; Yong Li; Wen Tang; Buyun Shi; Lijun Wang; Xiaolan Shu; Ying Jiang; Chenguang Qin; Hui Xu
Journal:  Brain Behav       Date:  2021-09-14       Impact factor: 2.708

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.