Steam sterilization of laparoscopic instruments.

C R Voyles1, D L Sanders, J E Simons, E A McVey, W B Wilson.   

Abstract

Because of the intricate internal parts of laparoscopic instruments, questions have been raised about the efficacy of cleaning and sterilization techniques. To assess these risks, hamburger meat was inoculated with high concentrations of vegetative pathogens and packed into laparoscopic cannulas. All openings of the cannulas were sealed during steam sterilization cycles ranging from 3 to 10 min in different experiments; cultures were obtained after cooling. Experiments were then performed using heat-resistant spore forms. Our studies showed that both the standard 10-min cycle and the 3-min "flash" were uniformly successful in killing all pathogenic microorganisms. A 7-min steam sterilization was necessary to kill spores within sealed cannulas. We conclude that a standard 10-min cycle within the steam autoclave provides complete sterilization of laparoscopic instruments; the 3-min "flash" sterilization is appropriate and safe for instruments that have been inadvertently contaminated or dropped during a surgical procedure.

Mesh:

Year:  1995        PMID: 7773462

Source DB:  PubMed          Journal:  Surg Laparosc Endosc        ISSN: 1051-7200


  2 in total

1.  Risk of Contamination in Assembled vs Disassembled Instruments in Hip Arthroplasty Surgery.

Authors:  Ryan R Mayer; S Samuel Bederman; Vincent M Colin; Martina M Berger; Thomas C Cesario; Ran Schwarzkopf
Journal:  J Arthroplasty       Date:  2016-02-09       Impact factor: 4.757

2.  Microbiological evaluation of the steam sterilization of assembled laparoscopic instruments.

Authors:  Tamara Carolina de Camargo; Kazuko Uchikawa Graziano; Alda Graciele Claudio Dos Santos Almeida; Karina Suzuki; Cely Barreto da Silva; Flávia Morais Gomes Pinto
Journal:  Rev Lat Am Enfermagem       Date:  2016-11-21
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.