Risk of Contamination in Assembled vs Disassembled Instruments in Hip Arthroplasty Surgery.

Ryan R Mayer1, S Samuel Bederman2, Vincent M Colin1, Martina M Berger3, Thomas C Cesario3, Ran Schwarzkopf4.   

Abstract

BACKGROUND: Periprosthetic joint infection (PJI) is one of the most common causes of revision total hip arthroplasty (THA) and associated with higher costs, prolonged pain, and worse clinical outcomes. Many factors have been linked to increased infection rates, one being the operative equipment and instrumentation used during the surgical procedure. With few arthroplasty instruments designed for complete disassembly and increasingly complex instrument designs, this study seeks to understand the effect that instrument disassembly plays on infection using disassembled and assembled standard femoral broach handles (BHs).
METHODS: Two BHs, not designed for disassembly, were modified and then contaminated in the disassembled state with Geobacillus stearothermophilus vegetative-form bacteria and spores. Using both flash and standard sterilization cycles, the BHs were steam sterilized in the disassembled or assembled state and then analyzed for remaining bacteria and spores.
RESULTS: At all target locations after either a flash sterilization cycle or a standard sterilization cycle, complete eradication of both the vegetative-form and spore-form of G stearothermophilus was achieved.
CONCLUSION: This study demonstrates that adequate decontamination of the tested BHs can be achieved after steam sterilization in either the disassembled or assembled state, without an increased risk of infection transmission.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  contamination; disassembled; instrumentation; periprosthetic joint infection; sterility; total hip arthroplasty

Mesh:

Year:  2016        PMID: 26948131      PMCID: PMC5922437          DOI: 10.1016/j.arth.2016.02.004

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  26 in total

1.  Use of light handles in the laminar flow operating theatre--is it a cause of bacterial concern?

Authors:  J R Hussein; R N Villar; A J Gray; M Farrington
Journal:  Ann R Coll Surg Engl       Date:  2001-09       Impact factor: 1.891

2.  Suction drain tip culture in orthopaedic surgery: a prospective study of 214 clean operations.

Authors:  B Sankar; P Ray; J Rai
Journal:  Int Orthop       Date:  2004-08-14       Impact factor: 3.075

3.  Epidemiology of surgical site infections after total hip and knee joint replacement during 2007-2009: a report from the VINCat Program.

Authors:  Joaquín López-Contreras; Enric Limón; Lourdes Matas; Montserrat Olona; Montserrat Sallés; Miquel Pujol
Journal:  Enferm Infecc Microbiol Clin       Date:  2012-06       Impact factor: 1.731

4.  Recommended practices for high-level disinfection.

Authors: 
Journal:  AORN J       Date:  2005-02       Impact factor: 0.676

5.  Gas and steam sterilization of assembled versus disassembled laparoscopic equipment. Microbiologic studies.

Authors:  P B Marshburn; W A Rutala; N S Wannamaker; J F Hulka
Journal:  J Reprod Med       Date:  1991-07       Impact factor: 0.142

6.  The impact of infection after total hip arthroplasty on hospital and surgeon resource utilization.

Authors:  Kevin J Bozic; Michael D Ries
Journal:  J Bone Joint Surg Am       Date:  2005-08       Impact factor: 5.284

7.  Steam sterilization of laparoscopic instruments.

Authors:  C R Voyles; D L Sanders; J E Simons; E A McVey; W B Wilson
Journal:  Surg Laparosc Endosc       Date:  1995-04

8.  Suction tip contamination in the ultraclean-air operating theatre.

Authors:  A H Robinson; S Drew; J Anderson; G Bentley; G L Ridgway
Journal:  Ann R Coll Surg Engl       Date:  1993-07       Impact factor: 1.891

9.  Periprosthetic joint infection: the incidence, timing, and predisposing factors.

Authors:  Luis Pulido; Elie Ghanem; Ashish Joshi; James J Purtill; Javad Parvizi
Journal:  Clin Orthop Relat Res       Date:  2008-04-18       Impact factor: 4.176

10.  Clinical outcome and patient satisfaction in aseptic and septic revision total knee arthroplasty.

Authors:  Ching-Jen Wang; Ming-Chun Hsieh; Ting-Wen Huang; Jun-Wen Wang; Han-Shiang Chen; Chen-Yeo Liu
Journal:  Knee       Date:  2004-02       Impact factor: 2.199

View more
  2 in total

1.  High bacterial contamination rate of electrocautery tips during total hip and knee arthroplasty.

Authors:  Hussein Abdelaziz; Akos Zahar; Christian Lausmann; Thorsten Gehrke; Helmut Fickenscher; Eduardo M Suero; Matthias Gebauer; Mustafa Citak
Journal:  Int Orthop       Date:  2018-02-12       Impact factor: 3.075

2.  Implant contamination as a cause of surgical site infection in spinal surgery: are single-use implants a reasonable solution? - a systematic review.

Authors:  Friederike Schömig; Carsten Perka; Matthias Pumberger; Rudolf Ascherl
Journal:  BMC Musculoskelet Disord       Date:  2020-09-25       Impact factor: 2.362

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.