Non-invasive estimation of venous admixture: validation of a new formula.

D A Hope1, B J Jenkins, N Willis, H Maddock, W W Mapleson.   

Abstract

We have developed a computer program that estimates venous admixture (intra-pulmonary shunt) from four measurements: haemoglobin concentration, end-tidal carbon dioxide tension (PE'CO2), fractional inspired oxygen concentration (FIO2) and pulse oximetry (SpO2). The formula was tested on patients in an intensive therapy unit by using it to estimate shunt while it was measured simultaneously by a standard, invasive method. A total of 101 measurements were made in 29 patients. After correcting the systematic errors in the assumed differences between PE'CO2 and arterial PCO2, and between SpO2 and co-oximetrically measured SaO2, and correcting for a trend in the arteriovenous oxygen concentration difference (C(a-v))2) with shunt, the bias of the non-invasive minus invasive shunt differences was negligible, with no significant dependence on shunt. The limits of agreement were then +/- 16% shunt overall (+/- 13% within patients). When SaO2 was used instead of SpO2, the limits were +/- 11% (+/- 8% within patients).

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7772428     DOI: 10.1093/bja/74.5.538

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  2 in total

1.  Non-invasive assessment of shunt and ventilation/perfusion ratio in neonates with pulmonary failure.

Authors:  H L Smith; J G Jones
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2001-09       Impact factor: 5.747

2.  Evaluation of a computer program for non-invasive determination of pulmonary shunt and ventilation-perfusion mismatch.

Authors:  Geoffrey G Lockwood; Nick L S Fung; J Gareth Jones
Journal:  J Clin Monit Comput       Date:  2014-01-09       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.