Factors affecting the in vivo precision of broad-band ultrasonic attenuation.

W D Evans1, E A Jones, G M Owen.   

Abstract

The in vivo precision of broad-band ultrasonic attenuation (BUA) was measured in one normal male volunteer over periods of one day and six weeks and in a group of 10 volunteers over a period of two weeks using a Walker Sonix UBA 1001 bone mineral analyser in which the foot is immersed in a water tank. Coefficients of variation ranged from 1-3% and, in general, there was no difference in BUA variance between right and left feet. However, right foot BUA values were consistently greater than those of the left, but there was no difference between mediolateral and lateromedial transmission. The use of preboiled water improved precision and reduced measurement time. In addition, factors affecting BUA precision were investigated in the individual volunteer. Equipment stability and conditions of immersion (immersion time, water depth, water temperature and concentration of a detergent wetting agent) had comparatively little effect on BUA values. On the other hand, the effects of foot positioning were more pronounced: rotation about the long axis of the foot, translation across the water tank and foot movement in the dorsal-plantar direction each contributed a maximum of 1.5-2.0% variation in BUA. However, both rotation about the long axis of the leg and foot movement in the heel-toe direction could change BUA by as much as 9%. It is likely that foot positioning will limit the precision achievable with this technique.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7708836     DOI: 10.1088/0031-9155/40/1/012

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  A nomogram for predicting osteoporosis risk based on age, weight and quantitative ultrasound measurement.

Authors:  C Pongchaiyakul; S Panichkul; T Songpatanasilp; T V Nguyen
Journal:  Osteoporos Int       Date:  2007-01-10       Impact factor: 4.507

2.  Assessment of the relationship between broadband ultrasound attenuation and bone mineral density at the calcaneus using BUA imaging and DXA.

Authors:  C Chappard; P Laugier; B Fournier; C Roux; G Berger
Journal:  Osteoporos Int       Date:  1997       Impact factor: 4.507

Review 3.  The role of ultrasound in the assessment of osteoporosis: a review.

Authors:  C F Njeh; C M Boivin; C M Langton
Journal:  Osteoporos Int       Date:  1997       Impact factor: 4.507

4.  The effect of ankle oedema on bone ultrasound assessment at the heel.

Authors:  A Johansen; M D Stone
Journal:  Osteoporos Int       Date:  1997       Impact factor: 4.507

5.  Effects of phase cancellation and receiver aperture size on broadband ultrasonic attenuation for trabecular bone in vitro.

Authors:  Jiqi Cheng; Frederick Serra-Hsu; Yuan Tian; Wei Lin; Yi-Xian Qin
Journal:  Ultrasound Med Biol       Date:  2011-10-26       Impact factor: 2.998

6.  Bone surface topology mapping and its role in trabecular bone quality assessment using scanning confocal ultrasound.

Authors:  Y Xia; W Lin; Y-X Qin
Journal:  Osteoporos Int       Date:  2007-03-15       Impact factor: 5.071

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.