Biomechanical comparison of the sliding hip screw and the dome plunger. Effects of material and fixation design.

J Choueka1, K J Koval, F J Kummer, G Crawford, J D Zuckerman.   

Abstract

We studied the biomechanical behaviour of three sliding fixation devices for trochanteric femoral fractures. These were a titanium alloy sideplate and lag screw, a titanium alloy sideplate and dome plunger with cement augmentation, and a stainless-steel sideplate and lag screw. We used 18 mildly osteoporotic cadaver femora, randomly assigned to one of the three fixation groups. Four displacement and two strain gauges were fixed to each specimen, and each femur was first tested intact (control), then as a two-part fracture and then as a four-part intertrochanteric fracture. A range of physiological loads was applied to determine load-bearing, load-sharing and head displacement. The four-part-fracture specimens were subsequently tested to failure to determine maximum fixation strengths and modes of failure. The dome-plunger group failed at a load 50% higher than that of the stainless-steel lag-screw group (p < 0.05) and at a load 20% higher than that of the titanium-alloy lag-screw group (NS). All 12 lag-screw specimens failed by cut-out through the femoral head or neck, but none of the dome-plunger group showed movement within the femoral head when tested to failure. Strain-gauge analysis showed that the dome plunger produced considerably less strain in the inferior neck and calcar region than either of the lag screws. Inferior displacement of the femoral head was greatest for the dome-plunger group, and was due to sliding of the plunger. The dome plunger with cement augmentation was able to support higher loads and did not fail by cut-out through the femoral head.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1995        PMID: 7706347

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  7 in total

1.  Biopolymer augmentation of the lag screw in the treatment of femoral neck fractures--a biomechanical in-vitro study.

Authors:  A Paech; E Wilde; A P Schulz; G Heinrichs; R Wendlandt; C Queitsch; B Kienast; Ch Jürgens
Journal:  Eur J Med Res       Date:  2010-04-08       Impact factor: 2.175

2.  One-year mortality rate after fragility hip fractures and associated risk in Nan, Thailand.

Authors:  Nuttorn Daraphongsataporn; Surapot Saloa; Kaiwan Sriruanthong; Nattaphon Philawuth; Krairoek Waiwattana; Prapan Chonyuen; Kununya Pimolbutr; Worapong Sucharitpongpan
Journal:  Osteoporos Sarcopenia       Date:  2020-06-23

3.  Cement Augmentation of Dynamic Hip Screw to Prevent Screw Cut Out in Osteoporotic Patients with Intertrochanteric Fractures: A Case Series.

Authors:  Avinash Kumar Rai; Rajesh Goel; Chirag Bhatia; Sumer Singh; Srikiran Thalanki; Ashwin Gondane
Journal:  Hip Pelvis       Date:  2018-12-06

4.  Validity of synthetic bone as a substitute for osteoporotic cadaveric femoral heads in mechanical testing: A biomechanical study.

Authors:  F O'Neill; F Condon; T McGloughlin; B Lenehan; C Coffey; M Walsh
Journal:  Bone Joint Res       Date:  2012-04-01       Impact factor: 5.853

5.  Analysis on the mechanical resistance of fixation of femoral neck fractures in synthetic bone, using the dynamic hip system and an anti-rotation screw.

Authors:  Anderson Freitas; Gustavo Melo Torres; André Cezar de Andrade de Mello E Souza; Rafael Almeida Maciel; Diogo Ranier de Macedo Souto; George Neri de Barros Ferreira
Journal:  Rev Bras Ortop       Date:  2014-10-27

6.  Biomechanical Comparison of Three Different Intramedullary Nails for Fixation of Unstable Basicervical Intertrochanteric Fractures of the Proximal Femur: Experimental Studies.

Authors:  Dae-Kyung Kwak; Won-Hyeon Kim; Sung-Jae Lee; Sang-Hyun Rhyu; Chul-Young Jang; Je-Hyun Yoo
Journal:  Biomed Res Int       Date:  2018-12-11       Impact factor: 3.411

7.  Proximal Femoral Nail Unlocked versus Locked (ProFNUL): a protocol for a multicentre, parallel-armed randomised controlled trial for the effect of femoral nail mode of lag screw locking and screw configuration in the treatment of intertrochanteric femur fractures.

Authors:  Arjun Sivakumar; Dominic Thewlis; Andreas Ladurner; Suzanne Edwards; Mark Rickman
Journal:  BMJ Open       Date:  2020-02-10       Impact factor: 2.692

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.