Twiddler's syndrome in children: an unusual cause of pacemaker failure.

S Abrams1, I Peart.   

Abstract

Mechanical causes of pacemaker failure are well recognised. Twiddler's syndrome leading to pacemaker failure has been previously recognised in adults, but there have been no published reports of its occurring in children. Two cases leading to failure of the pacing system are reported. In the first twiddling led to fracture of the lead and in the second it led to displacement of the lead from the heart. Children may be more susceptible to twiddler's syndrome because they have thinner subcutaneous tissues, making leads more accessible, and their comprehension of the consequences may be poor.

Entities:  

Mesh:

Year:  1995        PMID: 7696033      PMCID: PMC483790          DOI: 10.1136/hrt.73.2.190

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  6 in total

1.  Ventricular pacing from the atrial channel of a DDD pacemaker: a consequence of pacemaker twiddling?

Authors:  M H Anderson; A W Nathan
Journal:  Pacing Clin Electrophysiol       Date:  1990-12       Impact factor: 1.976

2.  Aggressive pacemaker twiddler's syndrome. Dislodgement of an active fixation ventricular pacing electrode.

Authors:  R B Lal; R D Avery
Journal:  Chest       Date:  1990-03       Impact factor: 9.410

3.  The pacemaker-twiddler's syndrome: a new complication of implantable transvenous pacemakers.

Authors:  C E Bayliss; D S Beanlands; R J Baird
Journal:  Can Med Assoc J       Date:  1968 Aug 24-31       Impact factor: 8.262

4.  The subpectoral pocket: the preferred implant site for pediatric pacemakers.

Authors:  P C Gillette; J Edgerton; J Kratz; V Zeigler
Journal:  Pacing Clin Electrophysiol       Date:  1991-07       Impact factor: 1.976

5.  Endocardial pacing in infants and children 15 kg or less in weight: medium-term follow-up.

Authors:  J A Till; S Jones; E Rowland; E A Shinebourne; D E Ward
Journal:  Pacing Clin Electrophysiol       Date:  1990-11       Impact factor: 1.976

6.  Pacing in children.

Authors:  H Ector; G Dhooghe; W Daenen; G Stalpaert; L van der Hauwaert; H de Geest
Journal:  Br Heart J       Date:  1985-05
  6 in total
  6 in total

1.  Special problems of pacing in children.

Authors:  Herwig Antretter; Joshua Colvin; Ulli Schweigmann; Herbert Hangler; Daniel Hofer; Karin Dunst; Josef Margreiter; Guenther Laufer
Journal:  Indian Pacing Electrophysiol J       Date:  2003-01-01

2.  Twiddler syndrome with a twist: a cause of vagal nerve stimulator lead fracture.

Authors:  Andrew T Trout; David B Larson; Francesco T Mangano; Cornelia H Gonsalves
Journal:  Pediatr Radiol       Date:  2013-07-07

3.  A rare cause of abdominal pain.

Authors:  Kathleen S Li; Haris A Khwaja; Tayyib T Hayat; Ayesha Asghar; Will Alsarakbi; Chris Kelley; Ekambaram D Babu
Journal:  Ann R Coll Surg Engl       Date:  2007-07       Impact factor: 1.891

4.  Twiddler's Syndrome: A Rare Hardware Complication in Spinal Cord Stimulation.

Authors:  Byung-Chul Son; Jin-Gyu Choi; Sang-Woo Ha
Journal:  Asian J Neurosurg       Date:  2018 Apr-Jun

5.  Use of an active fixation lead and a subpectoral pacemaker pocket may not avoid Twiddler's syndrome.

Authors:  Floris E A Udink Ten Cate; Roland Adelmann; Beate E Schmidt; Narayanswami Sreeram
Journal:  Ann Pediatr Cardiol       Date:  2012-07

6.  Wandering permanent pacemaker generators in children: a case series.

Authors:  Hilal Al Sabti; Raj Gopal Menon; Madan Mohan Maddali; John Valliattu
Journal:  J Med Case Rep       Date:  2008-05-18
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.