The reliability and accuracy of a standard method of tissue compliance assessment.

G Kawchuk1, W Herzog.   

Abstract

OBJECTIVE: To assess the reliability and the accuracy of a common method of tissue compliance measurement using a series of non-biological test surfaces (foam). Currently, tissue compliance measurement is most commonly obtained with a hand-held instrument known as a tissue compliance meter (TCM).
DESIGN: Descriptive study.
SETTING: Human Performance Laboratory, University of Calgary. INTERVENTION: A TCM was tested on four surfaces (three test and one control) with five different input forces resulting in 20 unique surface/force combinations. For each combination of surface/force, ten trials were obtained in a random order by each of five examiners, yielding a total of 1000 separate measurements of surface compliance. MAIN OUTCOME MEASURE: Millimeters of surface displacement per Newtons of input force.
RESULTS: Intraclass correlation coefficients were calculated for each of the 20 surface/force combinations to judge interexaminer reliability. The median of these 20 coefficients was 0.005: the greatest single value tending toward complete reliability was 0.22. Trials obtained from the control surface (assumed to be incompressible), demonstrated a range of displacements from 0.00 to 2.00 mm.
CONCLUSIONS: Within the design of this experiment, the reliability and accuracy of the hand-held tissue compliance meter was poor. We would suggest that the adequacy of this instrument in clinical practice or scientific work must be seriously questioned. Although the assessment of tissue compliance may be useful in the characterization of the musculoskeletal system and particularly in the assessment of treatment outcome, we surmise that a more reliable and accurate instrument is needed for the quantification of tissue compliance.

Entities:  

Mesh:

Year:  1995        PMID: 7673796

Source DB:  PubMed          Journal:  J Manipulative Physiol Ther        ISSN: 0161-4754            Impact factor:   1.437


  5 in total

1.  Reliability of the myotonometer for assessment of posterior shoulder tightness.

Authors:  Caitlyn M Kerins; Stephanie D Moore; Timothy A Butterfield; Patrick O McKeon; Timothy L Uhl
Journal:  Int J Sports Phys Ther       Date:  2013-06

2.  The examination of soft tissue compliance in the thoracic region for the development of a spinal manipulation training mannequin.

Authors:  David J Starmer; Sean A Duquette; Brynne E Stainsby; Anthony M Giuliano
Journal:  J Can Chiropr Assoc       Date:  2015-06

Review 3.  New techniques for the quantification of fibromyalgia and myofascial pain.

Authors:  Jeffrey R Basford; Kai-Nan An
Journal:  Curr Pain Headache Rep       Date:  2009-10

4.  Investigation of absolute intra-rater and inter-rater reliabilities during the muscle hardness estimation.

Authors:  Kazunori Morozumi; Katsuyuki Morishita; Masahiro Aoki; Yuka Yokoi; Takashi Inomata
Journal:  J Phys Ther Sci       Date:  2022-02-23

Review 5.  Content validity of manual spinal palpatory exams - A systematic review.

Authors:  Wadie I Najm; Michael A Seffinger; Shiraz I Mishra; Vivian M Dickerson; Alan Adams; Sibylle Reinsch; Linda S Murphy; Arnold F Goodman
Journal:  BMC Complement Altern Med       Date:  2003-05-07       Impact factor: 3.659

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.