Pulse oximetry in ventilated preterm newborns: reliability of detection of hyperoxaemia and hypoxaemia, and feasibility of alarm settings.

F Paky1, C M Koeck.   

Abstract

The aim of our study was to evaluate the reliability of pulse oximetry in detecting both hyper- and hypoxaemic states and to create clinically feasible alarm limits. A total of 792 readings of a pulse oximeter and corresponding values of arterial oxygen tension from 146 (79M, 67F) artificially ventilated preterm newborns with indwelling umbilical artery catheters were compared. Predictive value analysis of pulse oximeter readings related to arterial oxygen tension confirmed the ability of the pulse oximeter to identify both hypoxaemia and hyperoxaemia. However, a clinically feasible and safe range of alarm limits for maintenance of arterial oxygen tension of 40-90 mmHg (5.3-12 kPa) could only be established at a sensitivity level less than 0.9. At a level of 0.85, the alarm range on the pulse oximeter was 92.5-95%. Based on these findings, we are concerned about using pulse oximetry as the sole means of oxygen monitoring for preterm infants receiving supplementary oxygen. A combination of the pulse oximetry with other methods of blood-gas monitoring seems mandatory.

Entities:  

Mesh:

Year:  1995        PMID: 7670240     DOI: 10.1111/j.1651-2227.1995.tb13709.x

Source DB:  PubMed          Journal:  Acta Paediatr        ISSN: 0803-5253            Impact factor:   2.299


  5 in total

1.  Making ICU alarms meaningful: a comparison of traditional vs. trend-based algorithms.

Authors:  R Schoenberg; D Z Sands; C Safran
Journal:  Proc AMIA Symp       Date:  1999

2.  Optimal oxygen saturation in premature infants.

Authors:  Meayoung Chang
Journal:  Korean J Pediatr       Date:  2011-09-30

3.  Prophylactic nasal continuous positive airways pressure in newborns of 28-31 weeks gestation: multicentre randomised controlled clinical trial.

Authors:  F Sandri; G Ancora; A Lanzoni; P Tagliabue; M Colnaghi; M L Ventura; M Rinaldi; I Mondello; P Gancia; G P Salvioli; M Orzalesi; F Mosca
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-09       Impact factor: 5.747

4.  Detection of hyperoxaemia in neonates: data from three new pulse oximeters.

Authors:  B Bohnhorst; C S Peter; C F Poets
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-11       Impact factor: 5.747

5.  Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation?

Authors:  Gavin D Perkins; Daniel F McAuley; Simon Giles; Helen Routledge; Fang Gao
Journal:  Crit Care       Date:  2003-06-11       Impact factor: 9.097

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.