F S Nolte1, B Metchock, T Williams, L Diem, A Bressler, F C Tenover.
Abstract
We compared penicillin MICs obtained with three different commercially available broth microdilution panels (MicroScan, Sensititre, and Pasco) with MICs obtained with reference microdilution panels for 20 well-characterized pneumococci with decreased susceptibilities to penicillin (7 resistant and 13 intermediate). All panels were supplemented with 2 to 5% lysed horse blood (LHB) prepared in-house. Additional supplements included fastidious inoculum broth (FIB) for MicroScan panels and commercially prepared LHB (Difco) for Pasco panels. The percentages of penicillin-resistant strains (MIC 2 micrograms/ml) detected by the different methods follow: MicroScan-FIB, 0; MicroScan-LHB 0; Pasco in-house LHB, 71; and Sensititre-LHB, 100. The percentages of intermediate strains (MIC = 0.1 to 1.0 micrograms/ml) detected by the different methods follow: MicroScan-FIB, 31; MicroScan-LHB 23; Pasco in-house LHB, 46; and Sensititre-LHB, 85. Difco LHB supplement failed to support the growth of 86% of the strains in the Pasco panels. Of the commercially available panels evaluated, only Sensititre, supplemented with LHB prepared in-house could reliably detect penicillin-resistant pneumococci.Entities:
Mesh:
Substances:
Year: 1995 PMID: 7665649 PMCID: PMC228274 DOI: 10.1128/jcm.33.7.1804-1806.1995
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948