Literature DB >> 7656632

Evaluation of carbon dioxide rebreathing during pressure support ventilation with airway management system (BiPAP) devices.

F Lofaso1, L Brochard, D Touchard, T Hang, A Harf, D Isabey.   

Abstract

The purpose of this study was to evaluate whether carbon dioxide (CO2) rebreathing occurs in acute respiratory failure patients ventilated using the standard airway management system (BiPAP pressure support ventilator; Respironics; Murrysville, Pa) with positive inspiratory airway pressure and a minimal level of positive end-expiratory pressure (PEEP) and whether any CO2 rebreathing may be efficiently prevented by the addition of a nonrebreathing valve to the BiPAP system circuit. In the first part of the study, the standard device was tested on a lung model with a nonrebreathing valve (BiPAP-NRV) and with the usual Whisper Swivel connector (BiPAP-uc). With the BiPAP-uc device, the resident volume of expired air in the inspiratory circuit at the end of expiration (RVEA) was 55% of the tidal volume (VT) when the inspiratory pressure was 10 cm H2O and the frequency was at 15 cycles per minute. The BiPAP-NRV device efficiently prevented CO2 rebreathing but resulted in a slight decrease in VT, which was due to a significant increase in external PEEP (2.4 vs 1.3 cm H2O) caused by the additional expiratory valve resistance. For similar reasons, both the pressure swing necessary to trigger pressure support and the imposed expiratory work were increased in the lung model when the nonrebreathing valve was used. In the second part of the study, seven patients weaned from mechanical ventilation were investigated using a randomized crossover design to compare three situations: pressure support ventilation with a conventional intensive care ventilator (CIPS), BiPAP system use, and BiPAP-NRV. When we compared the BiPAP system use with the other two systems, we observed no significant effect on blood gases but found significant increases in VT, minute ventilation, and work of breathing. These findings are experimental and are clinical evidence that significant CO2 rebreathing occurs with the standard BiPAP system. This drawback can be overcome by using a non-rebreathing valve, but only at the expense of greater expiratory resistance.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7656632     DOI: 10.1378/chest.108.3.772

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  13 in total

Review 1.  Mechanical ventilation of the patient with severe chronic obstructive pulmonary disease.

Authors:  M T Gladwin; D J Pierson
Journal:  Intensive Care Med       Date:  1998-09       Impact factor: 17.440

2.  The configuration of bi-level ventilator circuits may affect compensation for non-intentional leaks during volume-targeted ventilation.

Authors:  Annalisa Carlucci; Annia Schreiber; Alessio Mattei; Alberto Malovini; Jacopo Bellinati; Piero Ceriana; Cesare Gregoretti
Journal:  Intensive Care Med       Date:  2012-09-28       Impact factor: 17.440

3.  Bilevel positive airway pressure ventilation: factors influencing carbon dioxide rebreathing.

Authors:  Zbigniew Szkulmowski; Kheirallah Belkhouja; Quoc-Hung Le; Dominique Robert; Laurent Argaud
Journal:  Intensive Care Med       Date:  2010-02-11       Impact factor: 17.440

4.  RIBRON: The spanish Online Bronchiectasis Registry. Characterization of the First 1912 Patients.

Authors:  Miguel Angel Martinez-García; Carmen Villa; Yadira Dobarganes; Rosa Girón; Luis Maíz; Marta García-Clemente; Oriol Sibila; Rafael Golpe; Juan Rodríguez; Esther Barreiro; Juan Luis Rodriguez; Rosario Menéndez; Concepción Prados; David de la Rosa; Casilda Olveira
Journal:  Arch Bronconeumol (Engl Ed)       Date:  2020-02-17       Impact factor: 4.872

5.  Effect of pressures and type of ventilation on aerosol delivery to chronic obstructive pulmonary disease patients.

Authors:  Marina E Boules; Nabila Ibrahim Laz; Ahmed A Elberry; Raghda R S Hussein; Mohamed E A Abdelrahim
Journal:  Beni Suef Univ J Basic Appl Sci       Date:  2022-04-15

Review 6.  Noninvasive positive pressure ventilation as treatment for acute respiratory failure in critically ill patients.

Authors:  M Antonelli; G Conti
Journal:  Crit Care       Date:  2000-01-24       Impact factor: 9.097

7.  Performance characteristics of seven bilevel mechanical ventilators in pressure-support mode with different cycling criteria: a comparative bench study.

Authors:  Yuqing Chen; Kewen Cheng; Xin Zhou
Journal:  Med Sci Monit       Date:  2015-01-26

8.  Evaluation of carbon dioxide rebreathing during exercise assisted by noninvasive ventilation with plateau exhalation valve.

Authors:  Yong-Er Ou; Zhi-Min Lin; Dong-Ming Hua; Ying Jiang; Ya-Ting Huo; Qun Luo; Rong-Chang Chen
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2017-01-16

9.  High flow biphasic positive airway pressure by helmet--effects on pressurization, tidal volume, carbon dioxide accumulation and noise exposure.

Authors:  Onnen Moerer; Peter Herrmann; José Hinz; Paolo Severgnini; Edoardo Calderini; Michael Quintel; Paolo Pelosi
Journal:  Crit Care       Date:  2009-06-05       Impact factor: 9.097

10.  Effect of a passive exhalation port on tracheostomy ventilation in amyotrophic lateral sclerosis patients: a randomized controlled trial.

Authors:  Andrea Vianello; Giovanna Arcaro; Beatrice Molena; Silvia Iovino; Federico Gallan; Cristian Turato; Rosario Marchese-Ragona
Journal:  J Thorac Dis       Date:  2018-02       Impact factor: 2.895