Literature DB >> 7651994

A filtered backprojection algorithm for pinhole SPECT with a displaced centre of rotation.

J Li1, R J Jaszczak, K L Greer, R E Coleman.   

Abstract

In this paper the importance of correcting a small centre-of-rotation displacement (approximately 1 mm) in single-photon-emission computed tomography (SPECT) using high-resolution pinhole collimation is demonstrated. A filtered backprojection (FBP) algorithm is derived for a pinhole geometry that has a displaced centre-of-rotation. The centre-of-rotation displacement, or mechanical shift (MS), is the displacement of the midplane of the pinhole collimator from the rotation centre. It is characterized by two orthogonal components: the shift eta of the midplane of the pinhole collimator along the direction of the axis of rotation, and the distance tau between the midline of the pinhole collimator and the axis of rotation. This algorithm is fast and corrects the centre-of-rotation displacement directly by incorporating this displacement into the algorithm. This new algorithm is evaluated using both a three-line source and a micro-SPECT cold rod phantom. The results demonstrate that the pinhole FBP with mechanical shift correction is able to correct the 'doughnut'-type artifacts caused by the mechanical shift and restore the expected system resolution.

Entities:  

Mesh:

Year:  1994        PMID: 7651994     DOI: 10.1088/0031-9155/39/1/010

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

Review 1.  Ultra-high-resolution imaging of small animals: implications for preclinical and research studies.

Authors:  D A Weber; M Ivanovic
Journal:  J Nucl Cardiol       Date:  1999 May-Jun       Impact factor: 5.952

2.  Evaluation of myocardial infarct size in rat heart by pinhole SPECT.

Authors:  T Hirai; R Nohara; R Hosokawa; M Tanaka; H Inada; Y Fujibayashi; M Fujita; J Konishi; S Sasayama
Journal:  J Nucl Cardiol       Date:  2000 Mar-Apr       Impact factor: 5.952

3.  Reproducibility of left ventricular volume and ejection fraction measurements in rat using pinhole gated SPECT.

Authors:  Christian Vanhove; Tony Lahoutte; Michel Defrise; Axel Bossuyt; Philippe R Franken
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-09-15       Impact factor: 9.236

4.  Body contour 180 degrees pinhole SPET with or without tilted detector: a phantom study.

Authors:  Alain Seret; David Flérès; Olivier Firket; Michel Defrise
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-06-17       Impact factor: 9.236

5.  A new reconstruction strategy for image improvement in pinhole SPECT.

Authors:  Tsutomu Zeniya; Hiroshi Watabe; Toshiyuki Aoi; Kyeong Min Kim; Noboru Teramoto; Takuya Hayashi; Antti Sohlberg; Hiroyuki Kudo; Hidehiro Iida
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-03-17       Impact factor: 9.236

Review 6.  Advances in pinhole and multi-pinhole collimators for single photon emission computed tomography imaging.

Authors:  Jalil Pirayesh Islamian; AhmadReza Azazrm; Babak Mahmoudian; Esmail Gharapapagh
Journal:  World J Nucl Med       Date:  2015 Jan-Apr
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.