Literature DB >> 7641371

Repeat dilation of intravascular stents in congenital heart defects.

F F Ing1, R G Grifka, M R Nihill, C E Mullins.   

Abstract

BACKGROUND: Intravascular (Palmaz) stents have been successfully implanted in patients with congenital and acquired branch pulmonary stenosis. Early results are excellent; however, there is no information on restenosis and repeat dilation in patients with congenital heart disease. The purpose of this study is to review the incidence of restenosis and demonstrate the safety and efficacy of repeat dilation of stents in this group of patients. METHODS AND
RESULTS: Of 94 patients with 163 implanted stents in this single-center study, 43 patients with 73 implanted stents underwent recatheterization. Only 2 of 73 restudied stents (3%) developed significant restenosis. In 20 patients, 30 stents were redilated. At stent implantation, the mean age of this subgroup was 14.2 years, the mean intraluminal diameter increased from 4.9 to 10.7 mm (P = .0001), and the systolic gradient (mean) across the stent decreased from 52 to 11 mm Hg (P = .0001). At recatheterization (mean, 13 months), all stents were patent. The mean diameter decreased by 1.2 mm (P = .0001), but the increase in the gradient (mean, 3 mm Hg) was not significant (P = .11). After repeat dilation, the diameter increased from 9.5 to 12.2 mm (P = .0001), and the gradient decreased from 14 to 8 mm Hg (P = .0003). The 2 stents with restenosis were redilated successfully. Two patients underwent a successful second redilation of 3 stents at 18 and 26 months. There were no complications.
CONCLUSIONS: All stents remained patent. The occurrence of significant restenosis is low (3%), and these restenoses can be redilated and/or restented. Repeat dilation of the Palmaz stent implanted in branch pulmonary artery stenosis can be performed with safety and efficacy (94% success rate) up to 3 years after stent implantation.

Entities:  

Mesh:

Year:  1995        PMID: 7641371     DOI: 10.1161/01.cir.92.4.893

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  27 in total

1.  Three new applications of stent technology in a single patient.

Authors:  C J McMahon; H G El-Said; C E Mullins
Journal:  Tex Heart Inst J       Date:  2001

2.  A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal-results 6-18 months after implantation into New Zealand white rabbits.

Authors:  M Peuster; P Wohlsein; M Brügmann; M Ehlerding; K Seidler; C Fink; H Brauer; A Fischer; G Hausdorf
Journal:  Heart       Date:  2001-11       Impact factor: 5.994

3.  Vessel restenosis versus "streamlining".

Authors:  C E Mullins
Journal:  Heart       Date:  2002-11       Impact factor: 5.994

Review 4.  Contributions of the Texas Children's Hospital Pediatric Cardiology Program to the field of pediatric cardiology.

Authors:  J T Bricker
Journal:  Tex Heart Inst J       Date:  1997

5.  Interventional cardiology.

Authors:  K P Walsh
Journal:  Arch Dis Child       Date:  1997-01       Impact factor: 3.791

6.  Balloon pulmonary valvuloplasty and stent implantation. For peripheral pulmonary artery stenosis in Alagille syndrome.

Authors:  A S Saidi; J P Kovalchin; D J Fisher; G D Ferry; R G Grifka
Journal:  Tex Heart Inst J       Date:  1998

7.  Stent Therapy for Aortic Coarctation in Children <30 kg: Use of the Low Profile Valeo Stent.

Authors:  Sok-Leng Kang; Andrew Tometzki; Demetris Taliotis; Robin Martin
Journal:  Pediatr Cardiol       Date:  2017-08-05       Impact factor: 1.655

8.  The Use and Outcomes of Small, Medium and Large Premounted Stents in Pediatric and Congenital Heart Disease.

Authors:  Brian A Boe; Jeffrey D Zampi; Kurt R Schumacher; Sunkyung Yu; Aimee K Armstrong
Journal:  Pediatr Cardiol       Date:  2016-08-27       Impact factor: 1.655

9.  The efficacy and safety of stent redilatation in congenital heart disease.

Authors:  C Duke; E Rosenthal; S A Qureshi
Journal:  Heart       Date:  2003-08       Impact factor: 5.994

10.  The role of stents in the treatment of congenital heart disease: Current status and future perspectives.

Authors:  Bjoern Peters; Peter Ewert; Felix Berger
Journal:  Ann Pediatr Cardiol       Date:  2009-01
View more