Evaluation of the i-STAT system: a portable chemistry analyzer for the measurement of sodium, potassium, chloride, urea, glucose, and hematocrit.

T Mock1, D Morrison, R Yatscoff.   

Abstract

OBJECTIVE: To evaluate the analytical performance of the i-STAT system, which is designed for point of care testing and employs a hand-held chemistry analyzer and disposable cartridges, which in the configuration tested, are capable of measuring sodium, potassium, chloride, urea, glucose, and hematocrit in approximately 65 microL of blood in 90's.
METHODS: Linearity and imprecision in hematocrit measurement were assessed using whole blood, while that for the other analytes were evaluated with aqueous solutions. The accuracy of the i-STAT system was judged by assay of patient specimens obtained both by venipuncture and fingerprick and correlated with the Kodak Ektachem 700XR and the microhematocrit methods.
RESULTS: Linearity was obtained over the clinically relevant range for all analytes. Total imprecision as expressed by the coefficient of variation (CV) was less than 3.5% for all analytes at both high and low concentration except for a low concentration of urea where a CV of 9.4% was obtained. Linear regression analysis revealed minimal systematic errors. The standard error about the regression line (Sy/x) ranged from 0.017 for hematocrit to 2.262 for chloride in the assay of venous blood, whereas in the assay of capillary blood the Sy/x ranged from 0.018 for hematocrit to 0.755 for glucose.
CONCLUSIONS: The analytical performance of the i-STAT was deemed acceptable by calculation of total error and comparison with published performance standards. Our study has shown the i-STAT system to be reliable, robust, and simple to operate. Moreover, the compactness of the analyzer and the requirement for only small volumes of whole blood will make it a valuable diagnostic tool in point of care settings.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7628079     DOI: 10.1016/0009-9120(95)90708-6

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  7 in total

1.  Hemodilution is not critical in the pathogenesis of the acute coagulopathy of trauma.

Authors:  Max Valentin Wohlauer; Ernest E Moore; Nathan M Droz; Jeffrey Harr; Eduardo Gonzalez; Miguel Fragoso; Christopher C Silliman
Journal:  J Surg Res       Date:  2011-05-19       Impact factor: 2.192

2.  Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients.

Authors:  Jacoline Steinfelder-Visscher; Steven Teerenstra; Jacqueline M T Klein Gunnewiek; Patrick W Weerwind
Journal:  J Extra Corpor Technol       Date:  2008-03

3.  Evaluation of the i-STAT (portable clinical analyser) for measuring haematological parameters in Atlantic cod (Gadus morhua) at different CO2 and temperature conditions.

Authors:  Radoslav S Borissov; Sølvi Espeland; Martin H Iversen
Journal:  Fish Physiol Biochem       Date:  2019-10-03       Impact factor: 2.794

Review 4.  Multianalyte Physiological Microanalytical Devices.

Authors:  Anna Nix Davis; Adam R Travis; Dusty R Miller; David E Cliffel
Journal:  Annu Rev Anal Chem (Palo Alto Calif)       Date:  2017-06-12       Impact factor: 12.400

5.  Analysis of mathematical modelling on potentiometric biosensors.

Authors:  N Mehala; L Rajendran
Journal:  ISRN Biochem       Date:  2014-05-07

6.  Reliability of the i-STAT for the determination of blood electrolyte (K+, Na+, and CI-) concentrations in cattle.

Authors:  E Yildirim; T Karapinar; A Hayirli
Journal:  J Vet Intern Med       Date:  2015-01       Impact factor: 3.333

7.  Cold Storage Exacerbates Renal and Mitochondrial Dysfunction Following Transplantation.

Authors:  S Shrum; L A MacMillan-Crow; N Parajuli
Journal:  J Kidney       Date:  2016-02-27
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.