Literature DB >> 7586262

Steroid elution improves the stimulation threshold in an active-fixation atrial permanent pacing lead. A randomized, controlled study. Model 4068 Investigators.

G H Crossley1, J A Brinker, D Reynolds, W Spencer, W B Johnson, H Hurd, L Tonder, M Zmijewski.   

Abstract

BACKGROUND: Prior work suggests that the addition of a steroid-eluting reservoir to a passive-fixation permanent pacemaker lead improves the stimulation threshold; however, no large randomized study has addressed this tissue. Over the last several years, there has been an increase in enthusiasm for the use of active-fixation permanent pacemaker leads for various reasons in spite of the generally accepted notion that active-fixation leads have higher stimulation thresholds. METHODS AND
RESULTS: This multicenter, randomized, controlled study examined the difference in performance between a standard active-fixation atrial lead (Medtronic model 4058) and a steroid-eluting lead (Medtronic model 4068). Stimulation thresholds were obtained in a four-point strength-duration fashion. Evaluations of sensing and impedance were performed as well. These evaluations were performed at implantation, at weeks 1 through 4, and at weeks 6, 12, 24, and 52. Stimulation thresholds were significantly better in the steroid lead than in the nonsteroid lead at each measurement point from 1 week to 12 months. The mean 1.6-V stimulation threshold at 12 months was 0.19 +/- 0.2 ms in the steroid lead and 0.41 +/- 0.30 ms in the control lead. No acute peaking was observed with the steroid lead, whereas significant peaking was observed with the control lead. There was no difference in long-term sensing or impedance.
CONCLUSIONS: Inclusion of a steroid-eluting reservoir in an active-fixation permanent pacing lead improved stimulation thresholds in both the subacute and chronic periods and therefore should extend pulse-generator longevity.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7586262     DOI: 10.1161/01.cir.92.10.2935

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  11 in total

1.  Reduced acute inflammatory responses to microgel conformal coatings.

Authors:  Amanda W Bridges; Neetu Singh; Kellie L Burns; Julia E Babensee; L Andrew Lyon; Andrés J García
Journal:  Biomaterials       Date:  2008-09-19       Impact factor: 12.479

2.  A Materials Roadmap to Functional Neural Interface Design.

Authors:  Steven M Wellman; James R Eles; Kip A Ludwig; John P Seymour; Nicholas J Michelson; William E McFadden; Alberto L Vazquez; Takashi D Y Kozai
Journal:  Adv Funct Mater       Date:  2017-07-19       Impact factor: 18.808

3.  Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict.

Authors:  M C Post; W Budts; A Van de Bruaene; R Willems; B Meyns; F Rega; M Gewillig
Journal:  Neth Heart J       Date:  2011-08       Impact factor: 2.380

4.  Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients.

Authors:  Elizabeth B Fortescue; Charles I Berul; Frank Cecchin; Edward P Walsh; John K Triedman; Mark E Alexander
Journal:  J Interv Card Electrophysiol       Date:  2005-10       Impact factor: 1.900

Review 5.  Engineering immune-responsive biomaterials for skin regeneration.

Authors:  Pingli Wu; Yangyang Liang; Guoming Sun
Journal:  Biomater Transl       Date:  2021-03-28

6.  Permanent atrial pacing lead implant route after Fontan operation.

Authors:  Kazuhiro Takahashi; Frank Cecchin; Elizabeth Fortescue; Charles I Berul; Mark E Alexander; Edward P Walsh; Francis Fynn-Thompson; John K Triedman
Journal:  Pacing Clin Electrophysiol       Date:  2009-06       Impact factor: 1.976

Review 7.  Biomaterials: Foreign Bodies or Tuners for the Immune Response?

Authors:  Erminia Mariani; Gina Lisignoli; Rosa Maria Borzì; Lia Pulsatelli
Journal:  Int J Mol Sci       Date:  2019-02-01       Impact factor: 5.923

8.  Wireless, battery-free, fully implantable multimodal and multisite pacemakers for applications in small animal models.

Authors:  Philipp Gutruf; Rose T Yin; K Benjamin Lee; Jokubas Ausra; Jaclyn A Brennan; Yun Qiao; Zhaoqian Xie; Roberto Peralta; Olivia Talarico; Alejandro Murillo; Sheena W Chen; John P Leshock; Chad R Haney; Emily A Waters; Changxing Zhang; Haiwen Luan; Yonggang Huang; Gregory Trachiotis; Igor R Efimov; John A Rogers
Journal:  Nat Commun       Date:  2019-12-17       Impact factor: 17.694

Review 9.  Cardiovascular implantable electronic devices: a review of the dangers and difficulties in MR scanning and attempts to improve safety.

Authors:  Pei Ghim Poh; Charlene Liew; Colin Yeo; Le Roy Chong; Andrew Tan; Angeline Poh
Journal:  Insights Imaging       Date:  2017-06-17

10.  Pacemaker Implantation Associated Myocardial Micro-Damage: A Randomised Comparison between Active and Passive Fixation Leads.

Authors:  Patrick Blažek; Jerko Ferri-Certić; Hrvoje Vražić; Carsten Lennerz; Christian Grebmer; Kazuaki Kaitani; Martin Karch; Boris Starčević; Verena Semmler; Christof Kolb
Journal:  Sci Rep       Date:  2018-03-20       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.