Factors affecting total and "respirable" dose delivered by a salbutamol metered dose inhaler.

M L Everard1, S G Devadason, Q A Summers, P N Le Souëf.   

Abstract

BACKGROUND: Many factors contribute to the high variability of doses delivered to the lungs of patients using metered dose inhalers (MDIs). Relatively little attention has been paid to the contribution to this variability of the way in which the MDI is handled before the inhalation manoeuvre. Instruction leaflets often recommend procedures at odds with those used for in vitro testing of the device. The standard protocol for in vitro assessment of salbutamol MDIs involves shaking the MDI vigorously for 30 seconds and wasting the first two actuations. Subsequent actuations are introduced into the testing device at five second intervals. Patient instructions do not include a recommendation to waste the first two actuations and recommend a delay of one minute between actuations. A series of experiments was performed to determine whether such differences might be important.
METHODS: The total and "respirable" doses delivered by a salbutamol MDI (Ventolin, Allen & Hanburys) under various conditions were assessed with a multistage liquid impinger. The quantity of drug deposited on each stage was measured by an ultraviolet spectrophotometric method. The effect on the delivered dose of not shaking the canister, not wasting the first two doses, waiting 30 seconds between actuations, and using multiple rapid actuations was assessed by comparing the results with those obtained using the standard in vitro testing protocol.
RESULTS: Compared with a standard protocol, it was found that not shaking the MDI before use reduced the total and "respirable" dose by 25.5% and 35.7%, respectively. The dose delivered when actuating the MDI at 30 second intervals was no different from that when intervals was no different from that when intervals of five seconds were used. Two actuations separated by one second had no effect on the total dose but reduced the "respirable" dose by 15.8%, while four rapid actuations reduced the total and "respirable" doses by 8.2% and 18.2%, respectively. Storing the MDI stem down reduced the total and "respirable" dose delivered in the first actuation by 25.0% and 23.3% despite shaking the MDI before use.
CONCLUSIONS: MDIs containing drug in suspension must be shaken before use to resuspend the drug contained in the MDI, but shaking does not alter the composition of the suspension in the metering chamber and hence the dose in the first actuation remains low. Very rapid actuations can reduce the dose delivered per actuation, but salbutamol MDIs can be actuated immediately after a 10 second breath holding pause without affecting the dose delivered.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7570408      PMCID: PMC474646          DOI: 10.1136/thx.50.7.746

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  9 in total

1.  Pharmaceutical formulations--suspensions and solutions.

Authors:  P Edman
Journal:  J Aerosol Med       Date:  1994

2.  Drug delivery from holding chambers with attached facemask.

Authors:  M L Everard; A R Clark; A D Milner
Journal:  Arch Dis Child       Date:  1992-05       Impact factor: 3.791

3.  Effect of canister temperature on performance of metered-dose inhalers.

Authors:  A F Wilson; D S Mukai; J J Ahdout
Journal:  Am Rev Respir Dis       Date:  1991-05

4.  Low first-spray drug content in albuterol metered-dose inhalers.

Authors:  T D Cyr; S J Graham; K Y Li; E G Lovering
Journal:  Pharm Res       Date:  1991-05       Impact factor: 4.200

5.  The use of reservoir devices for the simultaneous delivery of two metered-dose aerosols.

Authors:  A R Clark; G Rachelefsky; P L Mason; M J Goldenhersh; A Hollingworth
Journal:  J Allergy Clin Immunol       Date:  1990-01       Impact factor: 10.793

6.  Cooling of metered-dose inhalers decreases pressure output from canisters.

Authors:  N B Hampson; M P Mueller
Journal:  N Engl J Med       Date:  1989-02-02       Impact factor: 91.245

7.  Multiple actuations of salbutamol MDI into a spacer device reduce the amount of drug recovered in the respirable range.

Authors:  P W Barry; C O'Callaghan
Journal:  Eur Respir J       Date:  1994-09       Impact factor: 16.671

Review 8.  Therapeutic aerosols 1--physical and practical considerations.

Authors:  S P Newman; S W Clarke
Journal:  Thorax       Date:  1983-12       Impact factor: 9.139

Review 9.  Clinical aerosols. I. Characterization of aerosols and their diagnostic uses.

Authors:  R V Lourenço; E Cotromanes
Journal:  Arch Intern Med       Date:  1982-11
  9 in total
  9 in total

1.  Wheeze in childhood: is the spacer good enough?

Authors:  Veena Rajkumar; Barathi Rajendra; Choon How How; Seng Bin Ang
Journal:  Singapore Med J       Date:  2014-11       Impact factor: 1.858

Review 2.  Antiasthmatic drug delivery in children.

Authors:  Elizabeth Biggart; Andrew Bush
Journal:  Paediatr Drugs       Date:  2002       Impact factor: 3.022

3.  Dose-response to salbutamol via a novel palm sized nebuliser (Aerodose Inhaler), conventional nebuliser (Pari LC Plus) and metered dose inhaler (Ventolin Evohaler) in moderate to severe asthmatics.

Authors:  Brian J Lipworth; Erika J Sims; Karla Taylor; Wendy Cockburn; Robert Fishman
Journal:  Br J Clin Pharmacol       Date:  2005-01       Impact factor: 4.335

4.  Controller Inhalers: Overview of Devices, Instructions for Use, Errors, and Interventions to Improve Technique.

Authors:  Patrick K Gleeson; Scott Feldman; Andrea J Apter
Journal:  J Allergy Clin Immunol Pract       Date:  2020-03-12

Review 5.  Use of inhaler devices in pediatric asthma.

Authors:  Fernando Maria De Benedictis; David Selvaggio
Journal:  Paediatr Drugs       Date:  2003       Impact factor: 3.022

6.  Variability in Delivered Dose from Pressurized Metered-Dose Inhaler Formulations Due to a Delay Between Shake and Fire.

Authors:  Ross H M Hatley; Jacob Parker; John N Pritchard; Dirk von Hollen
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2016-09-16       Impact factor: 2.849

Review 7.  Inhaler technique: facts and fantasies. A view from the Aerosol Drug Management Improvement Team (ADMIT).

Authors:  Mark L Levy; P N R Dekhuijzen; P J Barnes; M Broeders; C J Corrigan; B L Chawes; L Corbetta; J C Dubus; Th Hausen; F Lavorini; N Roche; J Sanchis; Omar S Usmani; J Viejo; W Vincken; Th Voshaar; G K Crompton; Soren Pedersen
Journal:  NPJ Prim Care Respir Med       Date:  2016-04-21       Impact factor: 2.871

8.  Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study.

Authors:  Roberto W Dal Negro; Pietro Longo; Orestepaolo Villanis Ziani; Luca Bonadiman; Paola Turco
Journal:  Multidiscip Respir Med       Date:  2017-05-29

9.  Errors in Aerosol Inhaler Use and Their Effects on Maternal and Fetal Outcomes among Pregnant Asthmatic Women (Subanalysis from QAKCOP Study).

Authors:  Wanis H Ibrahim; Fatima Rasul; Mushtaq Ahmad; Abeer S Bajwa; Laith I Alamlih; Anam M El Arabi; Dhabia Al-Mohannadi; Mohammed Y Siddiqui; Israa S Al-Sheikh; Azdin A Ibrahim
Journal:  Can Respir J       Date:  2018-12-18       Impact factor: 2.409

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.