Single- and double- lumen silicone breast implant integrity: prospective evaluation of MR and US criteria.

W A Berg1, C I Caskey, U M Hamper, J E Kuhlman, N D Anderson, B W Chang, S Sheth, E A Zerhouni.   

Abstract

PURPOSE: To evaluate the accuracy of magnetic resonance (MR) and ultrasound (US) criteria for breast implant integrity.
MATERIALS AND METHODS: One hundred twenty-two single-lumen silicone breast implants and 22 bilumen implants were evaluated with surface coil MR imaging and US and surgically removed. MR criteria for implant failure were a collapsed implant shell ("linguine sign"), foci of silicone outside the shell ("noose sign"), and extracapsular gel, US criteria were collapsed shell, low-level echoes within the gel, and "snowstorm" echoes of extracapsular silicone.
RESULTS: Among single-lumen implants, MR imaging depicted 39 of 40 ruptures, 14 of 28 with minimal leakage; 49 of 54 intact implants were correctly interpreted. US depicted 26 of 40 ruptured implants, four of 28 with minimal leakage, and 30 of 54 intact implants. Among bilumen implants, MR imaging depicted four of five implants with rupture of both lumina and nine of 10 as intact; US depicted one rupture and helped identify two of 10 as intact. Mammography accurately depicted the status of 29 of 30 bilumen implants with MR imaging correlation.
CONCLUSION: MR imaging depicts implant integrity more accurately than US; neither method reliably depicts minimal leakage with shell collapse. Mammography is useful in screening bilumen implant integrity.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7568852     DOI: 10.1148/radiology.197.1.7568852

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  Study of breast implant rupture: MRI versus surgical findings.

Authors:  A Vestito; F F Mangieri; A Ancona; C Minervini; V Perchinunno; S Rinaldi
Journal:  Radiol Med       Date:  2012-03-19       Impact factor: 3.469

Review 2.  The effect of study design biases on the diagnostic accuracy of magnetic resonance imaging for detecting silicone breast implant ruptures: a meta-analysis.

Authors:  Jae W Song; Hyungjin Myra Kim; Lillian T Bellfi; Kevin C Chung
Journal:  Plast Reconstr Surg       Date:  2011-03       Impact factor: 4.730

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.