Influence of internal current and pacing current on pacemaker longevity.

A Schuchert1, K H Kuck.   

Abstract

UNLABELLED: The effects of lower pulse amplitude on battery current and pacemaker longevity were studied comparing the new, small-sized VVI pacemaker, Minix 8341, with the former model, Pasys 8329. Battery current was telemetrically measured at 0.8, 1.6, 2.5, and 5.0 V pulse amplitude and 0.05, 0.25, 0.5, and 1.0 msec pulse duration. Internal current was assumed to be equal to the battery current at 0.8 V and 0.05 msec. Pacing current was calculated subtracting internal current from battery current. The Minix pacemaker had a significantly lower battery current because of a lower internal current (Minix: 4.1 +/- 0.1 microA; Pasys: 16.1 +/- 0.1 microA); pacing current of both units was similar. At 0.5 msec pulse duration, the programming from 5.0-2.5 V pulse amplitude resulted in a greater relative reduction of battery current in the newer pacemaker (51% vs 25%). Projected longevity of each pacemaker was 7.9 years at 5.0 V and 0.5 msec. The programming from 5.0-2.5 V extended the projected longevity by 2.3 years (Pasys) and by 7.1 years (Minix). The longevity was negligibly longer after programming to 1.6 V.
CONCLUSION: extension of pacemaker longevity can be achieved with the programming to 2.5 V or less if the connected pacemakers need a low internal current for their circuitry.

Entities:  

Mesh:

Year:  1994        PMID: 7511227     DOI: 10.1111/j.1540-8159.1994.tb01344.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Feasibility of using the automatic generating system for quartz watches as a leadless pacemaker power source.

Authors:  H Goto; T Sugiura; Y Harada; T Kazui
Journal:  Med Biol Eng Comput       Date:  1999-05       Impact factor: 2.602

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.