The device developed by the authors and described here enables the user to measure the depth from the water surface to the point of measurement for a cylindrical ion chamber with a waterproof plastic cap in a water phantom, free of surface-tension error with a high precision. The device seeks vertical orientation and provides the convenience of hands-free operation. The measurement process is simple and quick with a precision of 0.1 mm. (The device is currently available as a 'water phantom depth gauge' from Nuclear Associates, Division of Victoreen Inc., Clare Place, NY, USA.)

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7480120     DOI: 10.1088/0031-9155/40/8/007