Evaluation of absorbed dose in mammography: monte carlo simulation studies.

K Doi, H P Chan.   

Abstract

Computer programs for calculation of the backscatter factor and absorbed dose in breast phantoms are developed based on Monte Carlo simulation. Rad/roentgen conversion factors are calculated for water, fat, a mixture of 50% water and 50% fat, and lucite irradiated with monoenergetic and polyenergetic x rays from a tungsten or molybdenum anode x-ray tube. These factors can be used to estimate skin dose, average depth dose, or average integral dose in a breast when the beam quality, exposure in air at the skin position, and composition of the breast are known. Calculated backscatter factors are considerably greater than measured values reported previously.

Entities:  

Mesh:

Year:  1980        PMID: 7360961     DOI: 10.1148/radiology.135.1.7360961

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  Radiation dose to organs and tissues from mammography: Monte Carlo and phantom study.

Authors:  Ioannis Sechopoulos; Sankararaman Suryanarayanan; Srinivasan Vedantham; Carl J D'Orsi; Andrew Karellas
Journal:  Radiology       Date:  2007-12-04       Impact factor: 11.105

2.  Characterization of the homogeneous tissue mixture approximation in breast imaging dosimetry.

Authors:  Ioannis Sechopoulos; Kristina Bliznakova; Xulei Qin; Baowei Fei; Steve Si Jia Feng
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

  2 in total