Literature DB >> 7358570

Intraocular lens optical quality: update 1979.

R J Olson.   

Abstract

One hundred and twenty lenses from 12 manufacturers were randomly selected for optical studies using an optical bench. Resolution efficiency was measured with an Air Force target, and measurements of back focal length in air were used to calculate back focal length in aqueous. There were no lenses with: less than 50% resolution efficiency, more than a 0.5-D variation from the stated power, or more than 0.25 D of toricity. Seventy-five percent of the lenses had at least 74% resolution efficiency; 93% were within 0.25 D of the stated power; 95% had 0.1 D or less toricity. Lenses of recent manufacture exhibit a marked improvement in optical quality. Optical parameters for all lenses observed in this study were within good clinical ranges.

Mesh:

Year:  1980        PMID: 7358570     DOI: 10.1016/s0146-2776(80)80099-1

Source DB:  PubMed          Journal:  J Am Intraocul Implant Soc        ISSN: 0146-2776


  2 in total

1.  Accuracy of routine intraocular lens power calculation in a district general hospital.

Authors:  P T Gregory; R M Esbester; D L Boase
Journal:  Br J Ophthalmol       Date:  1989-01       Impact factor: 4.638

2.  Mechanical and biochemical effects of man-made fibres and metals in the human eye, a SEM-study.

Authors:  W L Jongebloed; M J Figueras; D Humalda; L J Blanksma; J G Worst
Journal:  Doc Ophthalmol       Date:  1986-01-15       Impact factor: 2.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.