Corrections for the response time and delay of mass spectrometers.

R Arieli, H D Van Liew.   

Abstract

Computations based on mass spectrometer (MS) outputs will be incorrect unless the delay for drawing the sample into the instrument and response time of the instrument are accounted for. When we changed concentration abruptly in two different mass spectrometers, the responses were sigmoid shaped not exponential, and time constants derived from the main part of the response curves were 43-60 ms; single-exponent corrections using these values caused the corrected waveform to overshoot. For a better correction, we used a two-exponent correction, C2 = Co + (Y1 + Y2) (d2C0/dt) + Y1Y2 (d2C0/dt2), where C0 is MS output as a function of time t, C2 is corrected concentration, and Y1 and Y2 are time constants. Assumption of a third exponent was of little value. For a successful correction Y1 must be smaller than a measured one-exponent time constant. We used two-thirds of the measured value for Y1 and then calculated Y2 from the once-corrected response. The second-order correction approximates a square output in response to a square input. To deal with delay time in a way that would give good resolution of dynamic changes and also be compatible with our response-time correction, we corrected for the difference between time of the input event and time that the output reaches 20% of full response. We validated our methods by integrations of amounts of gases drawn into and out of a syringe and in human breaths.

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 7319875     DOI: 10.1152/jappl.1981.51.6.1417

Source DB:  PubMed          Journal:  J Appl Physiol Respir Environ Exerc Physiol        ISSN: 0161-7567


  11 in total

1.  A real-time algorithm to improve the response time of a clinical multigas analyser.

Authors:  L Wong; R Hamilton; E Palayiwa; C Hahn
Journal:  J Clin Monit Comput       Date:  1998-08       Impact factor: 2.502

2.  Transmural gradient of tissue gas tensions in the canine left ventricular myocardium during coronary clamping and reactive hyperemia.

Authors:  B I Levy; E Pinard; J B Michel; A Tedgui; J Seylaz
Journal:  Pflugers Arch       Date:  1986-10       Impact factor: 3.657

3.  Automatisation of the single-breath nitrogen washout test.

Authors:  M F Melo; A Giannella Neto
Journal:  Med Biol Eng Comput       Date:  1989-03       Impact factor: 2.602

4.  Response characteristics of a mass spectrometer.

Authors:  S J Tavener; S J Withy; E A Harris
Journal:  Med Biol Eng Comput       Date:  1984-11       Impact factor: 2.602

5.  Lack of agreement between gas exchange variables measured by two metabolic systems.

Authors:  Djordje G Jakovljevic; David Nunan; Gay Donovan; Lynette D Hodges; Gavin R H Sandercock; David A Brodie
Journal:  J Sports Sci Med       Date:  2008-03-01       Impact factor: 2.988

Review 6.  Algorithms, modelling and VO₂ kinetics.

Authors:  Carlo Capelli; Capelli Carlo; Michela Cautero; Cautero Michela; Silvia Pogliaghi; Pogliaghi Silvia
Journal:  Eur J Appl Physiol       Date:  2010-02-27       Impact factor: 3.078

Review 7.  Open-circuit respirometry: real-time, laboratory-based systems.

Authors:  Susan A Ward
Journal:  Eur J Appl Physiol       Date:  2018-05-04       Impact factor: 3.078

8.  Rate of change of alveolar carbon dioxide and the control of ventilation during exercise.

Authors:  C J Allen; N L Jones
Journal:  J Physiol       Date:  1984-10       Impact factor: 5.182

9.  An experimental set-up to test heat-moisture exchangers.

Authors:  N Unal; J C Pompe; W P Holland; I Gültuna; P E Huygen; K Jabaaij; C Ince; B Saygin; H A Bruining
Journal:  Intensive Care Med       Date:  1995-02       Impact factor: 17.440

10.  A realistic validation study of a new nitrogen multiple-breath washout system.

Authors:  Florian Singer; Birgitta Houltz; Philipp Latzin; Paul Robinson; Per Gustafsson
Journal:  PLoS One       Date:  2012-04-27       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.