Literature DB >> 722482

Assay of tocainide in blood by high-pressure liquid chromatography.

E M Wolshin, M H Cavanaugh, C V Manion, M B Meyer, E Milano, C R Reardon, S M Wolshin.   

Abstract

A sensitive, specific, high-pressure liquid chromatographic assay for the determination of tocainide in whole blood is described. The residue from a methylene chloride extract of alkalinized blood was resolvated in a mobile phase of methanol--water (47:53) containing 1% acetic acid and 6.16 mM 1-octanesulfonic acid, adjusted to pH 4.0. Chromatography was performed on a reversed-phase column with detection at 254 or 225 nm. The limits of accurate measurement were 2 microgram/ml for a 1-ml blood sample monitored at 254 nm and 0.2 microgram/ml for a 2-ml sample monitored at 225 nm. The assay was tested on samples from emergency protocol patients and was also found suitable for single-dose pharmacokinetic studies.

Entities:  

Mesh:

Substances:

Year:  1978        PMID: 722482     DOI: 10.1002/jps.2600671215

Source DB:  PubMed          Journal:  J Pharm Sci        ISSN: 0022-3549            Impact factor:   3.534


  2 in total

1.  High-performance liquid chromatographic analysis of tocainide in human plasma.

Authors:  L Conings; N Verbeke
Journal:  Pharm Res       Date:  1985-11       Impact factor: 4.200

Review 2.  Clinical pharmacokinetics of the newer antiarrhythmic agents.

Authors:  A M Gillis; R E Kates
Journal:  Clin Pharmacokinet       Date:  1984 Sep-Oct       Impact factor: 6.447

  2 in total