Literature DB >> 7166518

The fudged formula for intraocular lens power calculations.

H J Shammas.   

Abstract

The formula presented is a modification of Colenbrander's formula. The formula yields a 0.70 diopter stronger power emmetropizing lens for a 23.0 mm eye. It also fudges the axial length by subtracting 0.1 mm for each 1.0 mm longer than 23.0 mm, and by adding 0.1 mm for each 1.0 mm shorter than 23.0 mm. Axial length was measured using immersion A-scan technique. In our prospectively tested group, the adjusted formula predicted implant power within +/- 1.00 diopter in 95% of eyes.

Mesh:

Year:  1982        PMID: 7166518     DOI: 10.1016/s0146-2776(82)80028-1

Source DB:  PubMed          Journal:  J Am Intraocul Implant Soc        ISSN: 0146-2776


  2 in total

1.  Accuracy of routine intraocular lens power calculation in a district general hospital.

Authors:  P T Gregory; R M Esbester; D L Boase
Journal:  Br J Ophthalmol       Date:  1989-01       Impact factor: 4.638

2.  SRK II formula in the calculation of intraocular lens power.

Authors:  M S Dang; P P Raj
Journal:  Br J Ophthalmol       Date:  1989-10       Impact factor: 4.638

  2 in total