The Friedmann Visual Field Analyser Mark II--technical evaluation and clinical results.

R W de Boer, T J van den Berg, M R Beintema, E L Greve, J Höppener, W M Verduin.   

Abstract

The perimetric concept and the technical construction of the Friedmann Visual Field Analyser Mark II (VFA Mk II) were evaluated. This instrument for multiple stimulus static campimetry of the central 25 degrees is the successor of the Friedmann VFA Mk I. The number of stimuli is increased (99 vs. 47) and the console is improved. The positions of the presented stimuli are indicated on a LED-monitor. The mechanical and electrical construction of the VFA Mk II appeared to be good. 106 eyes of patients and normal subjects were tested and we conclude that the larger number of stimuli, though improving the chance of detecting small defects, complicates the manual registration and the interpretation of clinical results and prolongs the test, especially in the case of visual fields with substantial defects.

Entities:  

Mesh:

Year:  1982        PMID: 7166100     DOI: 10.1007/bf00216794

Source DB:  PubMed          Journal:  Doc Ophthalmol        ISSN: 0012-4486            Impact factor:   2.379


  4 in total

1.  The design of instrumentation for the efficient investigation of the visual fields.

Authors:  C H Bedwell
Journal:  Am J Optom Arch Am Acad Optom       Date:  1967-10

2.  Modification of the Friedmann visual field analyser (model I) to cover 104 points of measurement.

Authors:  W M Verduin; D Bakker
Journal:  Doc Ophthalmol       Date:  1982-01-29       Impact factor: 2.379

3.  The competer automatic campimeter I. Technical description and evaluation.

Authors:  R W de Boer; T J van den Berg
Journal:  Doc Ophthalmol       Date:  1982-12-01       Impact factor: 2.379

4.  Serial analysis of changes in visual field defects, employing a new instrument, to determine the activity of diseases involving the visual pathways.

Authors:  A I Friedmann
Journal:  Ophthalmologica       Date:  1966       Impact factor: 3.250

  4 in total
  3 in total

1.  The competer automatic campimeter I. Technical description and evaluation.

Authors:  R W de Boer; T J van den Berg
Journal:  Doc Ophthalmol       Date:  1982-12-01       Impact factor: 2.379

2.  Performance of computer assisted perimeters.

Authors:  E L Greve
Journal:  Doc Ophthalmol       Date:  1982-12-01       Impact factor: 2.379

3.  Evaluation of the Friedmann Visual Field Analyser Mark II. Part 1. Results from a normal population.

Authors:  D B Henson; S M Dix; A C Oborne
Journal:  Br J Ophthalmol       Date:  1984-07       Impact factor: 4.638