Literature DB >> 7156184

Backscatter factors for diagnostic radiology (1-4 mm Al HVL).

R M Harrison.   

Abstract

Backscatter factors (BSF) in the diagnostic radiology quality range (1-4 mm Al) have been determined by ionisation chamber and TLD techniques. Ionisation chamber measurements are subject to systematic errors due to displacement of the scattering medium, although high precision allows the demonstration of the kVp dependence of BSF for a given first HVL. TLD techniques lead to a reduction of these systematic errors although the read-out procedures lead to lower precision than ionisation chamber measurements. Comparison of the TLD-based measurements with the data of Br. J. Radiol. Supplement 11 indicates good agreement (less than 2%) for HVLS greater than 2 mm Al and field sizes greater than 10 cm x 10 cm. Comparison with the work of other authors not covered by Supplement 11 indicates, in general, good agreement only where kVp-filter combinations are similar. The importance is stressed of taking into account the kVp-filter combination when defining the quality of an X-ray beam.

Entities:  

Mesh:

Year:  1982        PMID: 7156184     DOI: 10.1088/0031-9155/27/12/005

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Calculation of the entrance skin dose distribution for fluoroscopically guided interventions using a pencil beam backscatter model.

Authors:  Sarath Vijayan; Zhenyu Xiong; Stephen Rudin; Daniel R Bednarek
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-14

2.  Fluoroscopic radiation exposure of the kyphoplasty patient.

Authors:  Bronek M Boszczyk; Michael Bierschneider; Stephanie Panzer; Werner Panzer; Roger Harstall; Katharina Schmid; Hans Jaksche
Journal:  Eur Spine J       Date:  2005-06-10       Impact factor: 3.134

3.  Optimisation of scoliosis examinations in children.

Authors:  Jolanta Hansen; Anne Grethe Jurik; Bente Fiirgaard; Niels Egund
Journal:  Pediatr Radiol       Date:  2003-09-05

4.  Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations.

Authors:  Gaetano Compagnone; Laura Pagan; Carlo Bergamini
Journal:  J Appl Clin Med Phys       Date:  2005-01-12       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.