In vitro determination of accuracy of cardiac output measurements by thermal dilution.

T V Bilfinger, C Y Lin, C E Anagnostopoulos.   

Abstract

The accuracy of cardiac output (C.O.) measurements by the thermodilution method was evaluated in an in vitro model within a flow range from 1 to 5 liters/min. For C.O. determinations, a 5F Swan-Ganz balloon-tipped thermodilution catheter and a 9520 Edwards computer were used. We made 420 measurements at known flow rates. In serial determinations, we achieved an overall accuracy of 86 to 93% compared to the reference flow; for single determinations, the accuracy ranged from 75 to 85%. The indicator volume (3, 5, or 10 cc) had no influence on the results. The thermodilution determinations at each flow rate were reproducible at between 2.5 and 8.5%. There was no difference in accuracy or reproducibility when ice-cold or room temperature saline was used. Caution in the interpretation of single C.O. determinations in low-flow states, i.e., in pediatric patients, is recommended.

Entities:  

Mesh:

Year:  1982        PMID: 7132327     DOI: 10.1016/0022-4804(82)90056-7

Source DB:  PubMed          Journal:  J Surg Res        ISSN: 0022-4804            Impact factor:   2.192


  5 in total

Review 1.  Cardiac output measurements. A review of current techniques and research.

Authors:  K C Ehlers; K C Mylrea; C K Waterson; J M Calkins
Journal:  Ann Biomed Eng       Date:  1986       Impact factor: 3.934

2.  Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.

Authors:  J D Mackenzie; N E Haites; J M Rawles
Journal:  Br Heart J       Date:  1986-01

3.  Investigation of a continuous heating/cooling technique for cardiac output measurement.

Authors:  K C Ehlers; K C Mylrea; J M Calkins
Journal:  Ann Biomed Eng       Date:  1987       Impact factor: 3.934

Review 4.  Errors in the measurement of cardiac output by thermodilution.

Authors:  T Nishikawa; S Dohi
Journal:  Can J Anaesth       Date:  1993-02       Impact factor: 5.063

5.  Comparison of bedside measurement of cardiac output with the thermodilution method and the Fick method in mechanically ventilated patients.

Authors:  Jésus Gonzalez; Christian Delafosse; Muriel Fartoukh; André Capderou; Christian Straus; Marc Zelter; Jean-Philippe Derenne; Thomas Similowski
Journal:  Crit Care       Date:  2002-12-20       Impact factor: 9.097