Literature DB >> 709807

Reliable fifteen-minute assay for theophylline.

K Nakatsu, J A Owen, K Scully.   

Abstract

We describe a reliable, rapid high pressure liquid chromatography assay for serum theophylline which can be performed on 0.5 ml or smaller samples. The procedure involves extraction, centrifugation and sampling all in the same disposable centrifuge tube. In the therapeutic range the coefficients of variation for within-day and day-to-day assays are 1.0% and 2.6%, respectively. The sensitivity of the method is better than 1 microgram/ml in serum. The minimum number of steps, rapid elution and use of only peak heights contribute to the reliability and rapidity of the assay.

Entities:  

Mesh:

Substances:

Year:  1978        PMID: 709807     DOI: 10.1016/s0009-9120(78)90332-6

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  2 in total

1.  Theophylline-7-acetic acid: lack of absorption and therapeutic effectiveness.

Authors:  J A Fleetham; J A Owen; B May; P W Munt; K Nakatsu
Journal:  Thorax       Date:  1979-08       Impact factor: 9.139

2.  Dose-dependency of theophylline clearance and protein binding.

Authors:  J A Fleetham; C E Bird; K Nakatsu; R D Wigle; P W Munt
Journal:  Thorax       Date:  1981-05       Impact factor: 9.139

  2 in total